3 MAINTENANCE
3.1
pH-meter scheduled maintenance
3.1.1 On-going inspection of the convertor unit, electrodes and connecting
cables for damage.
3.1.2 Cleaning of dirty external surfaces of the convertor unit using soft
detergents.
3.1.3 pH-meter calibration against buffer solutions according to 2.3.3.
pH-meter calibration against buffer solutions is to be undertaken:
once a month;
if pH-meter proper workability is doubtful;
after pH-meter repair or long storage; and
on electrode replacement.