35
On completion of measurements wash electrodes and the protective case
with distilled water and screw the base onto the case.
In case of long breaks in measurements remove the electrode (electrodes)
from the protective case and store them as described in their certificates.
To reduce electrolyte consumption in the electrode, the electrode filling hole
is to be closed within idling periods.
2.4.3 Procedure for measurements using FM-901/903 flow-through
module
To measure pH of "pure" water, use is to be made of the FM-901/903
ВР24.16.100 flow-through module from the pH-meter tools and accessories kit
supplied as agreed to by a customer.
The FM-901/903 flow-through module stabilizes water flow from a sampler,
ensures metering and visual control of water flow rate to the combined electrode. It
is intended to measure рН
25
in water with specific conductance over 0.2 μS/cm. At
conductance below 0.2 μS/cm measurement accuracy is not regulated.
pH is reduced to рН
25
in the range from plus 5 to plus 50.
To make measurements easy the FM-901/903 flow-through module and
convertor unit are placed on a support from MAPK-903 pH-meter tools and
accessories kit supplied as agreed to by a customer.
Add рН
25
measurement mode to the accessible modes list (
MEASURING
MODES
menu in 1.5.8.2).
Move from
MENU
mode to the measurement mode and select рН
25
measurement mode using the
key.
Measurements using FM-901/903 flow-through module are to be undertaken
as described in FM-901/903 ВР24.06.100РЭ flow-through module operation
manual.
MODE
▲