37
2.5.2 Setting of predetermined parameters of electrode system pH
calibration
Setting of predetermined parameters of electrode system pH calibration is
used for checking the transducer if pH-meter readings seem to be doubtful.
To set predetermined parameters, proceed as follows:
turn off pH-meter;
press the
key and holding it turn on the pH-meter;
depress the
key.
One of the displays shown in Fig. 2.11-2.13 shall come up.
Fig. 2.11
Fig. 2.12
Fig. 2.13
The display shall show computed values of electrode system parameters for
selected value of the isopotential point:
S
is electrode system slope;
pНi
and
Еi
are the coordinates of electrode system isopotential point versus
pНi value compliant with type of electrodes used and defined in
ADDITIONAL
SETTINGS
menu.
IMPORTANT: Once the pH-meter is de-energized and then turned on
the PROBE PARAMETERS menu shall show parameters of the pH-meter last
calibration rather than preset parameters!
RECORD
ENTER
RECORD
ENTER
INITIAL
pH PARAMETERS
ARE SET
S =
100 %
Ei =
00 mV
рНi =
7.00
INITIAL
pH PARAMETERS
ARE SET
S =
100 %
Ei =
0.0 mV
рНi =
4.00
INITIAL
pH PARAMETERS
ARE SET
S =
100 %
рНi =
6.7
Ei =
18 mV