31
4
If buffer solution pH value is not identified automatically, the display
shown in Fig. 2.3 shall come up. Refer to 2.5 of the Operation Manual
(Troubleshooting. Table 2.1).
Fig. 2.3
5
If buffer solution pH value is identified automatically, it shall appear and
the meter in the left part of the screen shall be pasted. Once readings
become stable, the meter shall be pasted, the display shown in Fig. 2.4
shall come up.
Fig. 2.4
6
Press the
key. The display shown in Fig. 2.5 shall come up. The pH-
meter shall enter the calibration mode against the second buffer solution
exhibiting pH = 9.18 at the solution temperature of (25.0±0.2) °С.
Fig. 2.5
RECORD
ENTER
WARNING!
CALIBRATION
ERROR!
CALIBRATION No 1
1.650
pH
23.0
С
А
CALIBRATION No 2
1.645 pH
PUT THE PROBE IN THE 2nd
BUFFER SOLUTION
AND PRESS
ENTER
Progress meter