34
Usually in measurements using electrodes readings stabilize within
maximum of 10 min. However, in some solutions at temperature close to 0
C
readings may stabilize within 15 min.
Storage between measurements is to meet requirements set out in the
electrode (electrodes) certificates. To reduce electrolyte consumption in the
electrode, the electrode filling hole is to be closed within idling periods.
2.4.2 Procedure of measurements using the protective case
The protective case protects electrodes in measurements and in pH-meter
haulage. It is supplied as agreed to by a customer with any type of electrodes
except ЭСКЛ-08М and ЭСКЛ-08М.1 combined laboratory glass electrodes.
Moisten an electrode (electrodes) and the sensor in distilled water and place
them in the protective case in the sockets as shown in Fig. 2.10.
Prior to measurements screw out the protective case base and proceed with
measurements as described in 2.4.1 immersing the case with electrodes and
temperature sensor into a solution.
Fig. 2.10
Protective case
Protective case base
Electrodes
Temperature sensor
Socks for electrodes
and temperature
sensor