SUI_I CR LF
- command comprehended but cannot be executed
MASS FRAME
- mass value in current unit is returned immediately
Frame format:
1
2
3
4
5
6
7-15
16
17 18 19
20 21
S
U
I
stabilit
y
space
sign
mass
space
unit
CR LF
Example:
S U I CR LF
– computer command
S U I ? _ - _ _ _ 5 8 . 2 3 7 _ k g _ CR LF
- command executed and mass returned
21.4.9. Switch on continuous transmission in basic unit
Syntax:
C1 CR LF
Possible answers:
C1_I CR LF
- command comprehended but cannot be executed
C1_A CR LF
MASS FRAME
-
command comprehended and in progress
-
mass value in basic unit is returned
Frame format:
1 2
3
4
5
6
7-15
16
17 18 19
20
21
S I
space
stability
space
sign
mass
space
unit
CR
LF
21.4.10.
Switch off continuous transmission in basic unit
Syntax:
C0 CR LF
Possible answers:
C0_I CR LF
- command comprehended but cannot be executed
C0_A CR LF
- command comprehended and executed
Содержание P2.H Series