21.4.4. Set tare value
Syntax:
UT_TARE CR LF
, where
TARE
– tare value
Possible answers:
UT_OK CR LF
- command executed
UT_I CR LF
- command comprehended but cannot be executed
ES CR LF
- command not recognized (possible wrong tare format)
Notice:
This protocol uses the dot character as a decimal point for separating the decimal fraction
part.
21.4.5. Send the stable result in basic unit
Syntax:
S CR LF
Possible answers:
S_A CR LF
S_E CR LF
-
command accepted and in progress
-
time limit for stable result exceeded
S_I CR LF
- command comprehended but cannot be executed
S_A CR LF
MASS FRAME
-
command accepted and in progress
-
mass value in basic unit is returned
Frame format:
1
2-3
4
5
6
7-15
16
17 18 19
20
21
S
space
stability
space
sign
mass
space
unit
CR
LF
Example:
S CR LF
– computer command
S _ A CR LF
- command accepted and in progress
S _ _ _ _ - _ _ _ _ _ _ 8 . 5 _ g _ _ CR LF
– command done, mass value in basic
unit is returned.
21.4.6. Send the result immediately in basic unit
Syntax:
SI CR LF
Possible answers:
SI_I CR LF
- command comprehended but cannot be executed at the moment