![VWR P2.H Series Скачать руководство пользователя страница 55](http://html1.mh-extra.com/html/vwr/p2-h-series/p2-h-series_user-manual_1057619055.webp)
21.2. A set of commands for RS interfaces
Commands
Description of commands
Z
Zeroing
T
Tarring
OT
Get tare
UT
Set tare
S
Send the stable result in basic unit
SI
Send the result immediately in basic unit
SU
Send the stable result in current unit
SUI
Send the result immediately in current unit
C1
Switch on continuous transmission in basic unit
C0
Switch off continuous transmission in basic unit
CU1
Switch on continuous transmission in current unit
CU0
Switch off continuous transmission in current unit
K1
Lock the scale keyboard
K0
Unlock the scale keyboard
NB
Give serial number
PC
Send all implemented commands
Notice:
1.
Each command have to be terminated in CR LF;
2.
The best Policy for communication is not sending another command until the
former answer has been received.
21.3. Respond message format
After sending a request message you can receive:
XX_A CR LF
command accepted and in progress
XX_D CR LF
command completed (appears only after XX_A)
XX_I CR LF
command comprehended but cannot be executed
XX _ ^ CR LF
command comprehended but time overflow error appeared
XX _ v CR LF
command comprehended but the indication below the
XX _ OK CR LF
Command done
ES_CR LF
Command not comprehended
XX _ E CR LF
error while executing command
– time limit for stable result exceeded
(limit time is a descriptive parameter of the scale)
XX
- command name
_
- substitutes spaces
Содержание P2.H Series
Страница 1: ...VWR P2 H Series USER MANUAL...
Страница 2: ...AUGUST 2017...
Страница 8: ...8 4 MAIN DIMENSIONS...
Страница 10: ...10 2 Remove the hole plug 3 Fix the hook 4 Turn the scale bottom side down...
Страница 64: ......