50
vwr.com I Instruction manual VWR
®
CO
2
incubator
Figure 54 - Sterilization completed
If during the heating up and holding phase:
the program is stopped by pressing the
button,
the external door is opened,
the unit is turned off with the main power switch or in case of power failure,
the unit’s error occurs,
the sterilization is prematurely terminated. The information of incompletely performed sterilization pops up on the display
and the event is registered in the Event log (“Sterilization failed”).
If the sterilization cycle was terminated prematurely, the germs inside the chamber might not be killed totally. It is
recommended to repeat the sterilization.
Figure 55 - Sterilization failed
If the door is open when the sterilization program is starting, the program won’t start. The information of cancelled sterilization
pops up on the controller display
Close the door and run the sterilization program again.
Figure 56 – Sterilization start cancelled