46
vwr.com I Instruction manual VWR
®
CO
2
incubator
Configuration of the sender e-mail account
In this window you can add the details of your e-mail account.
Server ID SMTP
Server password SMTP
Server Host SMTP
Server Port SMTP
You can also select the way of encryption
TLS
or
SSL
.
The above information can be obtained from the email account provider.
Connecting test
Confirms and saves changes
Cancels the entered changes
Corrections
10.14.
In this window
you can set correct temperature value indicated on the display by adding the correction value.
The set correction value is taken in the whole temperature range operation of the unit. For example, if the average temperature
displayed by the unit indicates 37°C and the average temperature measured by independent, external sensor indicates 37,5°C,
the correction should be set on +0,5°C. The average temperature should be calculated from chosen period of time e.g. 30min.
The correction range: -5°C to +5°C.
Before testing the connection, make sure that the unit is connected to a computer network
and has a properly configured network connection.
Figure 48 – E-mail - configuration of the sender e-mail account