18
vwr.com I Instruction manual VWR
®
CO
2
incubator
Information panel – four different windows, to scroll slide to the right or left (for more
information, see
Indicates which window is displayed
Information about the user currently logged in
Current date and time
Alarm bar
8.1.
The colour of the bar indicates the status of the unit:
– blue – the unit is working properly
– red – alarm
Alarms panel
8.2.
In this place appears the list of alarms. With active alarm, the control bar is red. The alarm event is displayed in the list, with
the status alarm active.
When an alarm event stops, the alarm remains in the list as inactive.
“
delete
” button – confirms and removes the alarm from the list (only inactive alarms can be deleted),
“
confirm
” button – confirms the class 3.1 alarm,
“
details
” button – displays a preview of all instances of selected alarm
.
If there are more alarms, a button
appears on the right side of the list that allows you to expand on the entire screen.