26
vwr.com I Instruction manual VWR
®
CO
2
incubator
Summary of segments
10.1.3.
In the summary in the figure all segments
are visible along with the entered settings:
segment number,
temperature, CO
2
concentration, duration time, target time of reaching temperature of a given segment.
With more parameters you can scroll the panel up and down.
Loop
10.1.4.
The option is available if the number of segments is equal to 2 or more.
When the program finishes the last segment, the unit starts the program again from the first segment.
You can define if a program should be carried out once (loop: 1) or multiple times (loop: 2 to 255). You can also set
the program to be carried out continuously, tick the „∞” option.
Priority
10.1.5.
Can be set in terms of:
Parameters
:
The program without a ramp – the unit starts the countdown of the segment time when the set temperature is reached.
The program with a ramp – first, the unit counts down the time of the ramp and then proceeds to the segment countdown when
the set temperature is reached. Regardless of whether the time of ramp elapsed.
Confirms and saves the changes
Cancels the entered changes to segments and goes into program parameters.
Figure 15 - The summary of the program