10 – VWR STandaRd Line PiPeTTORS 

instruction Manual – 11

5,000 and 10,000 μL pipettors

L +5mm

1.  Tip ejector disassembly 

 

Remove the tip ejector pushbutton.

 

Using a screwdriver, turn the metal arbor counter-

clockwise to unscrew the ejector from the arbor.

 

Release the ejector.

2.  Tip ejector assembly 

 

Remove the tip ejector pushbutton.

 

Using a screwdriver, turn the metal arbor clockwise 

to secure the ejector and adjust its length. 

NOTE: 

Leave a space of at least 1 mm between  

the shaft and tip ejector collar.

3.  Tip ejector adjustment 

 

Remove the ejector pushbutton.

 

Use the screwdriver to increase or decrease the 

ejector length.

 

Following ejector assembly or adjustment, ensure 

the ejector pushbutton is relocated.

 

if the tip ejector cannot be adjusted sufficiently, or if the 

diameter of the ejector is insufficient to eject the tip, it 

may be necessary to attach the ejector cap “

M

” to the 

ejector. 
in 2 and 10 μL pipettors place the cap, that can be 

purchased separately, on the bottom of the pipettor 

shaft and slide the cap upwards to the bottom of the 

tip ejector.

M

L

10000μL: L +

10mm

5000μL: L +6mm

M

Volume Setting

aspiration volume can be set by 

either turning the pipetting pushbut-

ton knob, or the adjustment knob. 

The volume setting is displayed as 

a three digit counter, which should 

be read from top to bottom. The 

smallest volume increment is printed 

on the bottom counter drum.

Examples of counter indications

The decimal point in 

the volume of aspi-

rated liquid (μL or mL) 

is indicated by change 

in color of digits. 

Black digits represent 

integer numbers, and 

red digits represent 

decimal fractions  

(up to 20 μL).

1

2
5

0

3
5

0

7
5

Volume Ranges

0.1-2 μL

5-50 μL

100-1,000 μL

Set Volume

1.25 μL

35 μL

750 μL (0.75 mL)

Increment

0.002 μL 

0.2 μL

2.0 μL

For maximum accuracy, the set volume must be 

approached from a higher value by decreasing counter 

readings. Before reaching the desired value, reduce the 

speed of turning the adjustment knob to avoid inadver-

tently exceeding the desired value. 

Содержание avantor

Страница 1: ...tandard Line Instruction Manual Pipettors VWR Standard Line VWR 8 channel Standard Line VWR 12 channel Standard Line This manual is available in additional languages at www vwr com PMS 647C CMYK 96 54...

Страница 2: ...s are designed for microplate filling The pipet tors enable precise and simultaneous delivery of 8 or 12 preset volumes of liquid These pipettors are available in four volume ranges 0 5 10 L 5 50 L 20...

Страница 3: ...ti channel manifold Contains a set of plungers and a set of flexible suspension shafts which func tion to reduce the force required to attach the tip Model Identification The volume range of the pipet...

Страница 4: ...Accuracy Precision Non filter Tips L Single channel Pipettors 2 Min 0 1 40 0 12 0 10 0 2 12 0 6 0 1 2 7 1 3 Max 2 1 5 0 7 10 Min 0 5 4 0 2 8 10 1 0 2 5 1 8 5 0 1 0 0 6 Max 10 0 0 5 0 4 20 Min 2 3 0 1...

Страница 5: ...o not attach the tips with a rocking movement as this may damage the shaft or plunger Observe this rule particularly with single channel pipettors of low volume range Never draw liquids directly into...

Страница 6: ...In 2 and 10 L pipettors place the cap that can be purchased separately on the bottom of the pipettor shaft and slide the cap upwards to the bottom of the tip ejector M L 10000 L L 10mm 5000 L L 6mm M...

Страница 7: ...ded to pre rinse the tip several times prior to use Do not pipette liquids with temperatures above 70 C NOTE When pipetting acids or corrosive solutions which emit vapors it is recommended to disassem...

Страница 8: ...ISO 8655 guidelines for the maximum nominal liquid volume drawn by the pipettor and for 10 of the maximum or minimum liquid volume according to the values given Section 5 The pipettors are designed t...

Страница 9: ...rm 10 aspirations and calculate the average value in mg Calculate the volume in L by multiplying the value in mg by the conversion factor Z L mg If the average aspirated volume exceeds the permissible...

Страница 10: ...accuracy should not alter if the pipet ting processes including autoclaving are carried out as described in this manual If a change in accuracy occurs it is recommended to Check the calibration of the...

Страница 11: ...ling elements are not sufficiently lubricated e g after repeated autoclaving procedures Incorrect aspiration Liquid with properties other than water den sity viscosity Calibrate the pipettor using the...

Страница 12: ...tional product information and special offers Contact us For information or technical assistance contact your local VWR representative or visit ww vwr com 14 Warranty VWR warrants that this product wi...

Страница 13: ...s Campus Northwest Business Park Ballycoolin Dublin 15 Tel 353 01 88 22 222 Email sales ie vwr com Italy VWR International S r l Via San Giusto 85 20153 Milano MI Tel 39 02 3320311 02 487791 Email inf...

Отзывы: