58
vwr.com I Instruction manual VWR® G Water System
f)
Click
Start
to enter cleaning process. Choose
Yes
on the pop-up dialog.
g)
System will be back to Standby mode after pH cleaning stops.
h)
Remove the Cleaning Pack and install P Pack back.
i)
Switch the system to Ready mode.
If the effect is not obvious after the cleaning, need to do pH cleaning once again. If there is still no
obvious improvement after cleaning twice, replace the RO membrane.