7. Adjust the position of the W lamp
Switch on the power,(the Switch Mirror should be placed to the position as indicates).
Observe the entrance facular, and it should in the center of the entrance hole. If the
facular deviate to Left or Right, then loosen the No.1 screws in Fig. 5-8 and move the
lamp seat to Left or Right until it focus on the center of the slot. Then fix the screws. If
the facular deviate to Up and Down, then loosen the No.2 screws and move the lamp
seat Up and Down until the facular focus on the center of the slot. Then fix the No. 2
screws again.