16
2) Set Wavelength. Press
to go into wavelength setting interface, press
,
to select measure method and press
to confirm. Input the
value of the wavelength and press
to confirm.
3) Blank. Put the Reference Sample in the light path and press
to do
blank.
4) Setup Standard Samples. Press
to setup standard, input the
concentrations of corresponding standard samples according the indication
and press
to confirm. Users can use
,
to choose the value
you just input and press
to delete, then input a new value, press
to confirm. Press
to cancel after all the input.
5) Calibrate Standard Samples. Put the corresponding standard samples in the
light path as the screen indicates and press
to measure. Then the
Abs. value would appear in the corresponding table.