Operating instructions, Quick installation
16
12. Read the transducer signal for calibration point 2 (read only).
Press
+
to get parameter ‘Transd.sign. p.2’ displayed.
Transd.sign. p.2
X.XXXXX mV/V
Make a note of the parameter value in the Quick set-up list, appendix 1.
13. Set the scale to zero.
This parameter displays the weight value with actual settings (but with one decimal
more than what is set in ‘Resolution’). Use it to set the weight value to ‘zero’ for
the unloaded scale.
Press
+
to get parameter ‘Set zero’ displayed, and check that the scale is unloaded.
Set zero
XXXX.XXX kg
Press ENTER to set the parameter value to zero.
Set zero
+0000.00
kg
The value is set to zero with leading sign and a cursor blinking.
Press ENTER for 2 seconds.
2 seconds
Set zero
0000.000 kg
The cursor disappears and the zero setting is accepted.
14. Read the zero offset.
By installation later of a replacement instrument, it is useful to know
the total zero offset value for the scale.
Press
+
to get parameter ‘Zero offset’ displayed.
Zero offset
X.XXX
kg
Make a note of the zero offset value in the Quick set-up list, appendix 1.
15. Exit ‘Quick set-up’.
Press ESCAPE to get to ‘Main menu Exit set-up’.
Main menu
Exit set-up
Содержание BHL Nobel AST 3
Страница 1: ......
Страница 2: ......
Страница 4: ......
Страница 24: ...Operating instructions Quick installation...
Страница 25: ......