Operating instructions, Quick installation
10
7. Edit the analogue output type.
The signal type for the analogue output is defined by parameter ‘Ana. output type’
that has a number of alternatives.
Press
+
to get parameter ‘Ana. output type’ displayed.
Ana. output type
4-20mA
Press
if the parameter should be edited, and make the editing
like in point
3.
on page 8.
8. Chose a calibration type.
Two calibration types can be chosen in ‘Quick set-up’: Data sheet and Deadweight.
As a new calibration shall be performed, AST 3P must first be set for editing,
indicated by a cursor at the parameter value. Then one of the alternatives must be
chosen and accepted.
Press
+
to get the parameter ‘Calibration type’ displayed.
Calibration type
Data sheet
The value points out which type of calibration that was performed last time.
If a new calibration shall be performed, press ENTER to set the instrument
for editing.
Calibration type
Data sheet
A cursor starts blinking to the left on the parameter value line.
Chose an alternative for the parameter like in point
3.
on page 8.
When
is pressed for 2 seconds, the chosen calibration type gets active and
the calibration can start.
The choice of calibration type determines which parameters will follow:
For Data sheet, se point
9.
on page 11.
For Deadweight, se point
9.
on page 14.