Intended use
Measuring and displaying electric parameters in the range of excess voltage cate-
gory III (up to max. 600V against ground potential, pursuant to EN 61010-1) and all
lower categories.
- Measuring direct voltages up to max. 600 V (CAT II)
- Measuring alternating voltages up to max. 600 V (CAT II)
- Measuring currents of up to max. 1000 A
- Measuring direct currents of up to max. 1000 A (VC-531 only)
- Frequency measuring up to 10 kHz
- Capacity measuring up to 4000 µF
- Measuring resistance values of up to 60 MOhm.
- Continuity check (< 70 Ohm acoustic)
- Diode test
The device may only be operated with the specified batteries.
The measuring instrument must not be operated when it is open, i.e. with an open
battery compartment or when the battery compartment cover is missing. Measuring
in damp rooms or under unfavourable ambient conditions is not admissible.
For safety reasons, when measuring only use measuring cables or accessories
which are adjusted to the specifications of the measuring device.
Unfavourable ambient conditions are:
- Wet conditions or high air humidity
- Dust and flammable gases, vapours or solvent,
- Thunderstorms or similar conditions such as strong electrostatic fields etc.
Any use other than that described above will damage the product and may involve
other risks, such as short-circuit, fire, electric shock, etc. Do not change or modify
any part of the product!
Please read the operating instructions carefully and keep them for later reference.
Please read the operating instructions thoroughly and keep the operating instruc-
tions for further reference.