- Press the “REC” button again to switch to display mode. Every keystroke now switches between the
display of “REC Max” and “REC Min”. The recording is halted in display mode.
- Press the “ESC” button (3) to start recording again. Min and max are no longer displayed.
- Hold down the “REC” button (6) for about 2 seconds to switch off the min/max memory function. The
“REC” symbol disappears.
Null balancing
The meter and the sensor can be null balanced. This becomes necessary when the meter shows a low
test result (instead of 0.0) in neutral position.
To null balance, proceed as follows:
- Connect the sensor to the meter and switch on the meter.
- Select acceleration “ACC” for measuring
- Leave the sensor in a motionless position. Wait until the display has stabilised.
- Hold down the “HOLD” button (3) for about 5 seconds. Null balancing concludes with a signal tone.
The display is set to zero.
☞
Null balancing has no influence on the measuring function, but it increases the accuracy of
measurements.
Null balancing is possible at zero deviation to a maximum of 10 digits (display shows e.g.
1.0, 0.10).
Data logger
The meter can record up to 500 results over a longer period of time. The automatic recording interval
can be set to 1 s, 2 s, 5 s, 10 s, 30 s, 60 s, 600 s, 1800 s, or 3600 s. You can also record manually (inter-
val: 0 s).
The setup menus to set and read out the parameters can only be exited by switching
off the meter.
Setting and adjusting the time interval
- Switch off the meter.
- Hold down buttons “HOLD” (3) and “REC” (7) and press the “POWER” button (4) to switch on the
device.
- Let go of all three buttons when you hear the beep. The current time interval is flashing. You are now
in programming mode.
- Press the “UNIT” button (6) to select the desired time interval. Every keystroke selects the next high-
est value. After the last value (3600) the display starts over (0).
0 = manual data logging
1 - 3600 = automatic data logging
- To confirm your selection, press the “REC” button (7) followed by the “ESC” button (3).
- Switch off the meter to exit the setup menu.