60
Frequency Measuring
range
Resolution Accuracy
De
θ
e
ESR /Rp
10 kHz
200.00
μ
H 0.01
μ
H ±(0,3%+2)
±0,003
±0,17°
±1.88L×10
2
+2
2000.0
μ
H 0.1
μ
H ±(0,3%+2)
±0,003
±0,17°
±1.88L×10
2
+2
20.000 mH
1
μ
H ±(0,3%+2)
±0,003
±0,17°
±1.88L×10
2
+2
200.00 mH
0.01 mH
±(0,3%+2)
±0,003
±0,17°
±1.88L×10
2
+2
2000.0 mH
0.1 mH
±(0,5%+3)
±0,005
±0,29°
±3.14L×10
2
+3
20.000 H
1 mH
±(2,0%+5)
±0,00
±1,15°
±1.26L×10
3
+5
Frequency Measuring
range
Resolution Accuracy
De
θ
e
ESR /Rp
100 kHz
20.000
μ
H 0.001
μ
H ±(0,5%+3)
±0,005
±0,29°
±3.14L×10
3
+3
200.00
μ
H 0.01
μ
H ±(0,5%+3)
±0,005
±0,29°
±3.14L×10
3
+3
2000.0
μ
H 0.1
μ
H ±(0,5%+3)
±0,005
±0,29°
±3.14L×10
3
+3
20.000 mH
1
μ
H ±(0,5%+3)
±0,005
±0,29°
±3.14L×10
3
+3
200.00 mH
0.01 mH
±(1,0%+5)
±0,010
±0,57°
±6.28L×10
3
+5
“L” corresponds to the value read in the unit Henry (H)
No voltages must be applied to the measuring inputs. Discharge all components
before connecting them to the measuring device. Do not touch any circuits or parts
of circuits if they may be subject to voltages higher than 25 V ACrms or 35 V DC!
Danger to life!