Specified measuring range: 5–100 % of the measuring range
600 V overload protection; Impedance: 400 kΩ (max. 250 V, 3 secs)
The multimeter may display ≤5 counts if a measuring input is short-circuited.
After using the LoZ feature, leave the multimeter for 1 minute before using it
again.
3
Alternating voltage (V/AC)
Range
Resolution
Accuracy
40.00 mV
0.01 mV
±(1.4% + 5)
400.0 mV
0.1 mV
4.000 V
0.001 V
±(1.3% + 4)
40.00 V
0.01 V
400.0 V
0.1 V
600 V
1 V
±(1.9% + 7)
Only available in "mV" mode
Specified measuring range: 5–100 % of the measuring range
Frequency range: 45–400 Hz; 600 V overload protection; Impedance: 10 MΩ
(mV: ≤100 MΩ)
The multimeter may display 5 counts if a measuring input is short-circuited.
TrueRMS peak (Crest Factor (CF)) ≤3 CF to 600 V
TrueRMS peak for non-sinusoidal signals plus tolerance
CF >1.0 - 2.0
+ 3%
CF >2.0 - 2.5
+ 5%
CF >2.5 - 3.0
+ 7%