35
15. tEChNICaL Data
general Information
Power supply .................................................2 x AAA batteries/rechargeable batteries
Operating time of the battery .........................approx. 50 h (without backlighting)
Deactivation function .....................................after 10 minutes
Measuring rate ...............................................10 x /s.
Measuring category .......................................CAT III 600 V
Pollution level ................................................2
Operating altitude ..........................................max. 3000 m
Opening range probe sensor .........................18 mm
Continuity Test ...............................................< 30 Ω
Input resistance .............................................1 MΩ (VDC/VAC)
Display ...........................................................2000 to 9999 digits
Display type ...................................................LC display with backlighting
PEAK hold recording .....................................< 10 ms (range: 10 - 282.8 A)
Max/Min hold recording .................................< 50 ms
Operating temperature range ........................-10 - +50 °C
Storage temperature range ...........................-30 - +60 °C
Dimensions ....................................................164 x 65 x 32 mm (L x W x H)
Weight ...........................................................175 g
measuring function
range and resolution
accuracy
AC current(50/60 Hz)
True RMS
200,0 A/AC
± (3.0 % + 10 digits)
DC current
200,0 A/DC
± (2.5 % + 8 digits)
DC voltage
600,0 V/DC
± (1.2 % + 4 digits)
AC voltage (50/60 Hz)
True RMS
600,0 V/AC
± (1.8 % + 10 digits)
Resistance
999,9 Ω
± (1.8 % + 8 digits)
The AC-V range is specific from 5 - 100 % of the measuring range.