S
ecti
on 5
- E
rro
r M
es
sages
Eccellenza Touch
VKI Publication #100308-003
81
Message Displayed
Code
Condition
Action(s) Required
Powder Dispensers current is under
minimum operation level
M1,
M2,
M3
Powder Dispenser current is below
the minimum operating level.
1 - Check the wiring and connections
to the powder dispenser motor.
2 - If message persists, replace the
powder dispenser motor.
Powder Dispensers Over Current
StandBy
Z99
The powder dispensers current sum
was over the configured stand by
operation level.
1 - Clear the message.
2 - Reboot the machine.
Powder Whippers Over Current
StandBy
Z99
Powder whippers current sum was
over the configured stand by
operation level
1 - Clear the message.
2 - Reboot the machine.
The water level is reaching the
safety level probes
L5
The water tank has overfilled and
the water level is reaching the safety
level probe.
1 - If this occurs during initial
heating, it may be due to water
expansion. Empty some water from
tank.
2 - The machine may have been
moved causing tank to overfill.
Empty some water from tank.
3 - Level probes require cleaning.
Valves & Relay current is over
safety level.
V1-V6
Valves & Relay current has exceeded
the maximum operating level.
1 - Clear the message.
2 - If message persists, the inlet
valve or one of the outlet valves
may need to be replaced.
Valves & Relay current is under
minimum operation level
V1-V6
Valves & Relay current is below the
minimum operating level.
1 - Clear the message.
2 - Check wiring and connections to
inlet and outlet valves.
3 - If message persists, the inlet
valve or one of the outlet valves
may need to be replaced.
Water Tank Coil Valves
Over Current StandBy
Z99
Valve(s) current has exceeded the
configured stand by operation level.
1 - Clears automatically. If not,
reboot the machine.
Water Tank Default
Critical Min Temp
H1
Water temperature went below the
configured minimum critical
temperature (Default 0 C).
1 - Clear the message.
2 - Reboot the machine.
Water Tank Default Fill
Heating Time
H1
Inlet valve and heater were both
activated for a period of time longer
than the configured timeout.
1 - Clear the message.
2 - Reboot the machine.
Water Tank Detectors
Power Failure
Z99
Fault signal triggered due to a failure
of the level probes and overflow
detector power supply.
1 - Clear the message.
2 - Reboot the machine.
3 - If message persists, replace the
control board.
Water Tank Heater
Over Current StandBy
Z99
Heater current has exceeded the
configured stand by operation level.
1 - Clears automatically. If not,
reboot the machine.
Water Tank Heating
Timeout
H1
Reaching the water set point
(Default 95 C) took longer than the
maximum configured heating
timeout (Default: 30 min.)
1 - Clear the message.
2 - Reboot the machine.
3 - If message persists, replace the
heater element.
Water Tank Heating Probe
Error
H1
Heater bad connection count has
exceeded the configured limit.
1 - Clear the message.
2 - Reboot the machine.
Water Tank Level Error
L4
Safety level probe is detected while
the normal level probe is not.
1 - Check wiring and connections to
the level probes.
2 - Clean level probes.
Water Tank Temperature
Probe Short
T1
Temperature probe is shorted
1 - Check the wiring and connection
to the temperature probe.
2 - Replace the temperature probe.
Water Tank Temperature
Probe Not Present
T1
Temp probe is open.
1 - Check the wiring and connection
to the temperature probe.
2 - Replace the temperature probe.