Vitalograph micro - Instructions for Use
Non-reproducible syringe strokes (red syringe
with ‘information’ in centre, after the 5th Stroke).
Outside 25%.
The following displays on power on if the previous calibration
verification has failed:
Calibration verification failed (orange syringe with
triangular warning symbol).
A Calibration Verification report may be printed or saved if required,
see section 5.3 Reporting.
If the procedure was followed correctly and the error icon is
showing, the calibration verification should be repeated. If the
error continues to show, contact Vitalograph using the contact
information at the start of this document.
Note:
To exit the Calibration Verification screen without performing
a check, press the forward key again to return to the Configuration
Menu screen. The Calibration Verification will not be logged to the
device memory in this case.
Note:
A calibration verification is recommended daily, if the device
has been stored or transported, if the flowhead is dropped or
replaced, and if EMI (Electromagnetic Interference) is suspected or
possible.
5.5.
Configuration Options
To access the Configuration menu, press the
icon on the main
screen.
There are four configuration options:
1. The Subject option
to configure: