Battery Depleted - an orange depleted battery
icon. Continued use is allowed, however prepare
to change the batteries, or connect via USB cable
to a computer to continue testing.
Battery Discharged - the Battery Discharged icon
displays on full screen on power up and a red
empty battery icon displays on the status bar.
Batteries should be changed immediately or the
device connected to a computer via USB cable to
continue testing.
When the USB power is connected, this icon
displays instead
of the battery icon
6.2. Power Save Mode
To improve battery life, when powered by batteries the micro will
auto power down if left unused for 5 minutes.
When running off the USB the device screen will go blank if left
unused for 5 minutes. The device will not auto power down if
powered by USB.
Pressing the screen or on/Off button will bring the device out
of power save mode.
7. Cleaning & Hygiene
7.1. Preventing Cross-Contamination of Subjects
A spirometer is not designed or supplied as a ‘sterile’ device.
Vitalograph intends that a new Bacterial Viral Filter (BVF) be used
for every subject to prevent cross contamination. Using a new
BVF provides a significant level of protection for the subject, the
device and the user against cross contamination during spirometry
manoeuvres.
The interior of a Vitalograph flowhead does not require
decontamination where a new BVF is used for each subject. The
outside surfaces of the device and flowhead tube may be cleaned
with a 70% isopropyl alcohol impregnated cloth to remove any
visible soiling and for low level disinfection.