26
Follow the on-screen instructions.
10. If a print out is required, press the
button.
Viewing/Printing the Accuracy Log
The accuracy check is added to a log. Follow these steps to view the
log.
1. Press the
More
button on the Home screen
2. Press the
Accuracy Check
button.
The log contains the following information:
a. Date of accuracy check
b. Time of accuracy check
c. Syringe serial number
d. Volume of syringe
e. Volume recorded by COMPACT.
The following *** will appear beside the log if the
calibration was updated on the unit.
It is possible to print the accuracy check log by
pressing the
button.
In addition to the above items, the following will be
shown on the printout:
f. User ID
g. Ambient temperature at time of accuracy check.
The date and time the log was printed and the
COMPACT serial number will be shown in the
headings.