27
When to Check Accuracy
•
In accordance with your own established procedures
•
After annual maintenance checks
•
After cleaning or disassembling spirometer for any reason
•
After adjusting calibration
•
If the flowhead has been dropped.
D
ELETING
S
UBJECT
I
NFORMATION AND
/
OR
T
EST
R
ESULTS
.
There are four choices available for deletion:
1. All tests and all subject information
2. All tests for all subjects but no subject information
3. All tests for a specific subject and his or her subject information
4. All tests for a specific subject but no subject information
Follow these steps to access information for deletion:
1. Press the
More
button on the Home screen
2. Press the
Database
button.
a. To delete all test and all subject information:
i.
Check the Select All Subjects check box.
ii.
Select the Delete Subject option.
iii.
Press the
Delete
button.
A confirmation message will appear
iv. To continue with the deletion, press the
Return
button.
v. To cancel the deletion, press the
Cancel
button.
b. To delete all tests for all subjects but no subject