“VitalPatch Off Body” Message

VitalPatch with Adhesive Overlay

If the screen displays the message “VitalPatch Off Body,” 

this indicates that the VitalPatch no longer detects 

proper skin contact. See figure to the right.
• 

If removal was intentional, tap 

Disconnect Patch

 to 

end the monitoring session.

• 

If removal was not intentional, inspect the VitalPatch 

for adhesive peel-off or lifting on the edges. If the 

adhesive looks intact, press the VitalPatch down 

against the skin to re-adhere and tap 

Wait for 

Reconnect

.

If the VitalPatch Reconnects

If reconnection after “VitalPatch Off Body” is successful, 

a “Posture Calibration Required” message will be 

displayed. Tap 

Calibrate

 to begin the posture 

recalibration process. See

 Step 5 and 6 beginning on 

page 11

.

If you do not recalibrate, VistaPoint will not provide you 

with respiration rate or posture while out of calibration. 

Accuracy of fall detection and automatic body 

temperature is also considered invalid. Heart rate and 

live ECG will continue to update continuously.

If the VitalPatch Does Not Reconnect

Replace the VitalPatch if:

• 

 The VitalPatch does not reconnect after several 

minutes.

•   The “VitalPatch Off” message continues.
• 

 The VitalPatch cannot be reapplied by pressing down 

to re-adhere.

•   The VitalPatch partially or completely falls off the skin.

If the VitalPatch must be replaced, select 

Disconnect 

Patch

 from the 

Menu

. See 

Section 10 on page 17

.

Using the Optional Adhesive Overlay  

to Prevent “VitalPatch Off Body”

If the VitalPatch will be exposed to conditions where 

lifting and patch-off may occur, you may apply an 

optional adhesive overlay to help keep the VitalPatch 

adhered to the skin. See 

Section 7 on page 10

.

Use only the approved adhesive overlay. The use of other 

adhesives or tape is not recommended and will prevent 

proper operation of the VitalPatch and will void the 

warranty of the VitalPatch. 

DO NOT

 use the adhesive 

overlay if the VitalPatch has already lifted off the skin. 

The overlay may be used if the VitalPatch begins to peel 

up at the edges but has not lost contact with the skin.

Section 19: 

“VitalPatch Off Body” Message

Waiting for VitalPatch 

to be reattached

VitalPatch b008bf_012345 Off Body

If the VitalPatch is okay, select “Wait 

for Reconnect”.

If the VitalPatch has been removed 

from the body, or is no longer needed, 

select “Disconnect Patch”

Wait for Reconnect

Disconnect Patch

Check that the VitalPatch biosensor is properly adhered to the skin.

Waiting for VitalPatch

Waiting for VitalPatch to be reattached to 

the skin.

More Information...

Overlay

VitalPatch

Skin

Содержание VistaPoint 3.2

Страница 1: ...vision History 2 Getting Started Setting Up VistaTablet or VistaPhone Device 3 Checking Device Battery Level and Patch Hours 5 Changing Patient Name and Device Name 6 Removing Previous VitalPatch 7 Ap...

Страница 2: ...ate spot check behavior streaming mode no longer supported for Nonin 3230 Rev E Added support VistaPoint Release 3 2 Changes include controls for modifying and displaying patient name and device name...

Страница 3: ...itoring session unpack the VistaTablet or VistaPhone device Keep the packaging that was provided Locate the following VistaTablet or VistaPhone USB charger and charging cable VitalPatch biosensors and...

Страница 4: ...d hold the Power Key until the Options menu displays and then select Power Off Unlocking the Device If the VistaTablet or VistaPhone device is asleep black screen unlock and start the VistaPoint appli...

Страница 5: ...ovider as soon as possible Failure to return may result in additional charges for monitoring services The monitoring equipment includes the following VistaTablet or VistaPhone USB charger and charging...

Страница 6: ...permission to change the patient name 3 Enter the new patient name in the space provided Allowed characters are letters numbers hyphens spaces periods and underscores No other characters are allowed...

Страница 7: ...r of the overlay and peel it off the skin b Continue pulling the overlay off Stretch the overlay horizontally to help release it from the skin Note The VitalPatch may begin pulling off with the overla...

Страница 8: ...inimum of 2 minutes before powering on the VitalPatch You may prepare the chest next step while waiting 2 Prepare Chest a If there is body hair on the site shave the site before cleaning Prepare an ar...

Страница 9: ...omfort or irritation occurs the VitalPatch and the optional adhesive overlay should be removed Contact your healthcare provider immediately If the VitalPatch begins to fall off or lift from the skin c...

Страница 10: ...erlay over the VitalPatch with the blue liners against the chest The raised portions of the VitalPatch should be aligned with the cutouts of the overlay See figure to the right b Ensure the tab you cr...

Страница 11: ...complete the help sequence Tap the Bluetooth ID of your VitalPatch ID when it appears on the screen If VistaPoint asks how you want to connect choose one of the following options Tap Search for VitalP...

Страница 12: ...g you to confirm your selection Note If you have a pacemaker VistaPoint will not display ECG heart rate or respiration rate If pacemaker status has been incorrectly selected it may be changed later Se...

Страница 13: ...ed after performing or skipping the temperature calibration See Section 9 on page 14 for more information 8 Recalibrate Body Temperature If your configuration supports automatic body temperature you m...

Страница 14: ...ed for a simplified display that does not show vital signs For these devices your vital signs continue to be collected and will be monitored remotely by your care team VitalPatch Menu 92 5 45 PM Repor...

Страница 15: ...a VitalPatch that does not support automatic body temperature selecting this option will display a unit setting screen See Section 13 on page 18 Flash Patch LED Use to check pairing between VistaPoin...

Страница 16: ...oaded automatically using approved Bluetooth devices Blood Oxygen Saturation as SpO2 Blood Pressure mmHg systolic and diastolic Weight pounds and kilograms Fall Detection If the VitalPatch detects a f...

Страница 17: ...l not be applied turn off the VistaTablet or VistaPhone by pressing and holding the Power Key on until the Options menu displays and then select Power Off CALL 911 if you are experiencing an emergency...

Страница 18: ...Measure temperature with a clinical grade thermometer 2 Tap the Menu button and then tap Calibrate Temperature or tap the Body Temperature tile on the dashboard 3 Tap the units Celsius C or Fahrenheit...

Страница 19: ...manual entry Section 15 Entering Body Temperature Manually Your care team may direct you to enter an oxygen saturation SpO2 reading manually on a regular schedule for example one or more times per day...

Страница 20: ...nter value and then re enter the value Section 17 Entering Blood Pressure Manually Tap Take BP or Take a Measurement Take Weight Manual Entry Use Bluetooth Weight value 206 6 Clear Values Enter Pounds...

Страница 21: ...ssage continues The VitalPatch cannot be reapplied by pressing down to re adhere The VitalPatch partially or completely falls off the skin If the VitalPatch must be replaced select Disconnect Patch fr...

Страница 22: ...ashboard will display a VitalPatch Not Detected message See Section 10 on page 17 for more information Message Posture Calibration Required If posture becomes Unknown VistaPoint will direct you to per...

Страница 23: ...t indicates that the VistaTablet or VistaPhone does not have a wireless connection or that the cloud service is down See messages above for more information Checking Your Wireless Connection Your Vist...

Страница 24: ...ing procedure 1 Plug VistaTablet into the charger If the battery is completely discharged the it may need several minutes to recharge before you can restart next step 2 Restart the VistaTablet by hold...

Страница 25: ...is ON Check the Bluetooth indicator in the display The indicator will be displayed when Bluetooth is ON If it is not ON use a long button press to turn it ON To power off the device press and hold th...

Страница 26: ...tion tile This will remove the pairing between VistaPoint and the Berry device Error Messages During Measurement The following error messages may be displayed in VistaPoint when taking a measurement S...

Страница 27: ...on again to move between options Long press the ON OFF button to select an option or to change its value Ensure Bluetooth is ON using the menu To power off the device press and hold the ON OFF button...

Страница 28: ...tion tile This will remove the pairing between VistaPoint and the Berry device Error Messages During Measurement The following error messages may be displayed in VistaPoint when taking a measurement S...

Страница 29: ...the device Release the device to close The display should face up b Check the display of the unit See figure to the right If the Bluetooth icon is displayed in the upper right of the display no additi...

Страница 30: ...he number under mac 3 Take a Reading in VistaPoint a Apply Masimo MightySat device to the finger The device will turn on and begin measuring See figure to the right b Tap Take SpO2 in the Oxygen Satur...

Страница 31: ...e team Note Monitor battery life and replace batteries as necessary 1 Find the Bluetooth ID Optional Note This step is optional and is only necessary when using the Model 3150 in an environment where...

Страница 32: ...finger h If the monitoring session is complete tap Disconnect in the Oxygen Saturation tile This will remove the pairing between VistaPoint and the Nonin 3150 Error Messages During Measurement The fo...

Страница 33: ...to the right The Bluetooth ID begins with BDA 2 Pair New in VistaPoint a Tap Take SpO2 in the Oxygen Saturation tile Tap Use Bluetooth and then tap Pair New Tap Nonin Model 3230 as the model b Apply t...

Страница 34: ...ading When the reading is complete the oxygen saturation will be displayed on the dashboard with an indicator of when the reading was taken If the oxygen saturation value is not displayed repeat the p...

Страница 35: ...y when using the UA 651 in an environment where multiple similar devices are being used such as a hospital To find the Bluetooth ID turn the base unit over and find the label above the battery compart...

Страница 36: ...on on the base unit to take a reading VistaPoint displays Please wait connecting while it retrieves the current reading When the reading is complete the blood pressure will be displayed on the dashboa...

Страница 37: ...o confirm the device pairing tap Confirm 3 Take a Reading a Tap Take Weight in the Weight tile b Tap Use Bluetooth if not already selected c Tap Take Reading or Take a Measurement on VistaPhone and th...

Страница 38: ...4 Airport Parkway Suite 300 San Jose CA 95110 408 963 4600 www vitalconnect com support vitalconnect com 2021 VitalConnect A list of our trademarks can be found at www vitalconnect com All other marks...

Отзывы: