VitalConnect VistaPoint 3.1 Скачать руководство пользователя страница 14

14

Temperature Calibration

Fahrenheit (°F)

Celsius (°C)

Take Temperature

Step 2

Take patient’s temperature using a clinical-grade 

thermometer. Enter value and select Calibrate.

Step 1

Select Units

Cancel

Calibrate

Section 14:

 Recalibrating Automatic Body Temperature

Confirm Recalibration

Select Units

Enter 

Temperature

When using the automatic body temperature feature, you 

should recalibrate body temperature each day between 

the hours of 09:00 and 21:00. The VistaPoint dashboard 

may also prompt you to recalibrate the body temperature 

as necessary.
To recalibrate the body temperature:
1.  Measure temperature with a clinical-grade 

thermometer.

2.  Tap the Menu button and then tap Calibrate 

Temperature (or tap the Body Temperature tile on the 

dashboard).

3.  Tap the units (Celsius °C or Fahrenheit °F).
4.  Enter the temperature and tap Calibrate. The 

VistaPoint dashboard will be displayed once 

calibration is complete.

If the pacemaker status has been entered incorrectly, 

enter a new status as follows:
1.  Tap the Menu button and then tap Pacemaker 

Settings.

2.  Select YES if you have a pacemaker. Select NO if you 

do not have a pacemaker. If you do not know, please 

ask your doctor.
After selecting a pacemaker status, a screen will 

display asking you to confirm your selection. 

Note: If you have a pacemaker, VistaPoint will not display 

ECG, heart rate or respiration rate. 

Menu

Calibrate Posture

Calibrate Temperature

Flash Patch LED

Disconnect Patch

Menu

Calibrate Posture

Calibrate Temperature

Flash Patch LED

Disconnect Patch

Confirm Status Change

Menu

Calibrate Posture

Calibrate Temperature

Flash Patch LED

Disconnect Patch

Section 15:

 Entering Body Temperature Manually

Section 16:

 Changing Pacemaker Status

When using manual entry of body temperature, enter a 

new body temperature reading as follows:
1.  Measure temperature with a clinical-grade 

thermometer.

2.  Tap the Body Temp tile in the dashboard (if using 

VistaPhone with simplified display, tap Take a 

Measurement button first).

3.  Tap the units (Celsius °C or Fahrenheit °F).
4.  Enter the temperature and tap OK. The VistaPoint 

dashboard will be updated with the new body 

temperature value and will display the time since the 

last manual entry.

Manual Entry

Body Temperature

--.-°

Manual Entry

Body Temperature

98.6°F

Just Now

Take Body Temperature

Fahrenheit (°F)

Celsius (°C)

Enter Temperature

Step 2

Use a clinical-grade thermometer to measure body

temperature. Enter value here and select Calibrate.

Step 1

Select Units

98.6

 

Enter

Select Units

Enter

Temperature

Select Pacemaker Status

Confirm Selection

Implanted Pacemaker Device

If you do not know, please ask your doctor.

Do you have an implanted pacemaker device?

YES, I have a pacemaker

NO, I do not have a pacemaker

Pacemaker Settings

Содержание VistaPoint 3.1

Страница 1: ...cal instructions The VistaPoint application collects and uploads vital signs from the VitalPatch and approved third party devices to your care team VistaPoint runs on the portable VistaTablet or Vista...

Страница 2: ...and enter the PIN provided by your care team default PIN is 1234 Section 1 Setting Up VistaTablet or VistaPhone Status Bar Status Bar Home Key Menu Menu Device Name Device Name VistaTablet Charging P...

Страница 3: ...the device will sound a warning tone and then turn off If this occurs charge the device before turning it on again Checking VitalPatch Hours Each VitalPatch will last up to 5 days 120 hours or 7 days...

Страница 4: ...oval pad underneath c Dispose the VitalPatch according to local laws for routine electronic waste d Clean skin with an alcohol pad and allow to dry Section 4 Removing Previous VitalPatch Sweep Adhesiv...

Страница 5: ...ace two fingers below the jugular notch dip at the base of the neck b Position diagonally with the three lines angled downward away from the midline VitalConnect Midline LEFT Side Left Hand Section 6...

Страница 6: ...back This creates a tab that will help you remove the top liner later See figure to the right 3 Place the Adhesive Overlay Over VitalPatch a Position the adhesive overlay over the VitalPatch with the...

Страница 7: ...Patch ID and then enter the last 6 characters of the Bluetooth ID 3 Choose Pacemaker Status After connecting the pacemaker status screen may appear The screen will appear only once per patient will no...

Страница 8: ...model If the Elevated option is not available ensure the patient is in the Standing Sitting or Supine position when calibrating Instruct patient to remain still in their current body position and tap...

Страница 9: ...lPatch Report Symptom Device Name Organization VitalPatch ID 0087ea has 94 hours remaining 92 5 45 PM Take SPO2 Take BP Take Weight Streaming Mode 98 Berry BM1000 00 01 0C 82 BF 6E Oxygen Saturation H...

Страница 10: ...contents will vary based on device configuration See figures to the right Disconnect Patch Use when monitoring is complete or to pair a new VitalPatch This option maintains the pairing with third part...

Страница 11: ...nt Discharge patient from the monitoring system Cancel Section 10 When Monitoring is Complete Tap Disconnect Patch Choose a Disconnect Option Section 11 Reporting a Symptom Section 12 Making a Video C...

Страница 12: ...e figure below Tap Calibrate to begin the posture recalibration process See Step 5 and 6 beginning on Page 7 If you do not recalibrate VistaPoint will not provide you with respiration rate or posture...

Страница 13: ...itionally if the displayed posture does not match the actual posture you must recalibrate the posture To do this select Posture Settings from the VitalPatch menu and then tap Calibrate Message Fall De...

Страница 14: ...ion Note If you have a pacemaker VistaPoint will not display ECG heart rate or respiration rate Menu Calibrate Posture Calibrate Temperature Flash Patch LED Disconnect Patch Menu Calibrate Posture Cal...

Страница 15: ...tton on VistaPhone and then Take BP 3 Tap Manual Entry if not already selected 4 Tap the first entry field Use the on screen keyboard to enter the two measured values and then tap Enter If you make a...

Страница 16: ...until the display turns off d After confirming operation proceed to Step 2 on page 17 for pairing Setting Up the BM2000A Wrist Refer to the BM2000A Operator s Manual for complete instructions See fig...

Страница 17: ...Saturation tile or Take a Measurement on VistaPhone and then Take SpO2 c Tap Use Bluetooth and then tap Pair New d Tap Berry BM2000A BM1000 as the model e Scanning for Devices will display Tap the Bl...

Страница 18: ...displayed in the upper right of the display no additional action is necessary go to Step 2 below c If Bluetooth is not ON press and hold the Touchpad until the menu screen is displayed Press and hold...

Страница 19: ...n VistaPhone and then Take SpO2 c Tap Use Bluetooth and tap Pair New d Tap Masimo MightySat as the model e Scanning for Devices will display Tap the Bluetooth ID of the Masimo MightySat when it appear...

Страница 20: ...our care team 1 Find the Bluetooth ID Optional Note This step is optional and is only necessary when using the Model 3150 in an environment where multiple similar devices are being used such as a hosp...

Страница 21: ...ings will continue to be transmitted to the dashboard as long as the Nonin 3150 is operational and the sensor is applied to the finger If the dashboard displays Finger Removed check finger is inserted...

Страница 22: ...will be measured and you will be returned to the dashboard If you see a message asking you to confirm the device pairing tap Confirm d Pairing is now complete Remove Model 3230 from finger 3 Take a Re...

Страница 23: ...us step b Tap Take BP in the Blood Pressure tile or Take a Measurement on VistaPhone and then Take BP c Tap Use Bluetooth d Tap Pair New e Tap Model UA 651 as the model f Scanning for Devices will dis...

Страница 24: ...ted c Tap Take Reading or Take a Measurement on VistaPhone and then Take Weight d Step on the scale Step off the scale when it beeps and displays Step Off VistaPoint displays Please wait connecting wh...

Страница 25: ...on 1 Tap the Menu button and then tap Applications menu item Select the Wi Fi Center application this application will not be displayed if your tablet uses a cellular connection 2 If Wi Fi is OFF tap...

Страница 26: ...ap the Volume Down button on the side of the VistaPhone The device will restart normally Bluetooth Connection Issues If you have problems connecting to or disconnecting from the VitalPatch the followi...

Отзывы: