Message: “VitalPatch Not Detected”

Indicates that the VitalPatch is out of range, has lost 

Bluetooth connection, or has expired.
•  If disconnection was intentional or VitalPatch has 

expired, tap Disconnect Patch to end the monitoring 

session.

•  If disconnection was not intentional, check that the 

VitalPatch is within Bluetooth connection range, and 

then tap Wait for Reconnect. Wait for up to one 

minute for the VitalPatch to automatically reconnect. 

When the VitalPatch returns within range of the tablet, 

it will automatically reconnect. 

This message may also be cleared by choosing 

Disconnect Patch and then repeating the connection 

process in 

Section 8 on page 7

.

Note: When the VitalPatch disconnects from the 

VistaTablet or VistaPhone, the VitalPatch will continue to 

save vital sign data for up to 10 hours. After 

reconnecting, this saved data will automatically be 

uploaded to the VistaTablet or VistaPhone. 

Message: “Patch Expires”

When the remaining hours on the VitalPatch are low, the 

dashboard will display “Patch Expires” warnings that 

show how many hours remain. Replace the VitalPatch 

before the expiration time. If the time limit is exceeded, 

the VitalPatch will be disconnected and the dashboard 

will display a “VitalPatch Not Detected” message. See 

Section 10 on page 11

 for more information. 

Message: Posture Calibration Required

If posture becomes “Unknown”, VistaPoint will direct you 

to perform the posture calibration procedure again. 

Additionally, if the displayed posture does not match the 

actual posture, you must recalibrate the posture. To do 

this, select Posture Settings from the VitalPatch menu 

and then tap Calibrate

Message: “Fall Detected”

If the VitalPatch detects a fall, VistaPoint displays a 

yellow “FALL Detected” alert on the dashboard. If this 

occurs, contact your care team.
To clear the fall notification, tap the Acknowledge button 

on the dashboard only when directed by your care team.

Message: Body Temperature Needs Calibration

If the body temperature displays “Needs Calibration”, 

you must perform a recalibration. See 

Section 14 on 

page 14

 

FALL

Detected

     Acknowledge

Calibration Required

The VitalPatch Biosensor requires 

calibration. Select “Calibrate” to begin the 

process.

Calibrate

Waiting for VitalPatch 

to return to connection 

range

WARNING: Patch Expires in 2 Hours

The VitalPatch Biosensor has less than 2 

hours of battery life remaining. Replace as 

soon as possible.

OK

VitalPatch Life 

is Low

VitalPatch b008bf_012345 Not Detected

If the Patch is Out of Range

If the Patch Has Been Removed

Make sure the VitalPatch is in range.
The VitalPatch biosensor will auto-
reconnect when back in range.

If the VitalPatch biosensor has been

removed from the body, or auto-

reconnect does not occur, select

“Disconnect Patch”.

Wait for Reconnect

Disconnect Patch

Waiting for Patch to Reconnect

Waiting for patient to return to connection 

range. The VitalPatch biosensor will auto-

reconnect when back in range.

More Information...

 

Posture

Unknown

Temperature Calibration Required

The VitalPatch Biosensor requires temperature 

calibration. Select “Calibrate” to begin

Calibrate

Not Now

Содержание VistaPoint 3.1

Страница 1: ...cal instructions The VistaPoint application collects and uploads vital signs from the VitalPatch and approved third party devices to your care team VistaPoint runs on the portable VistaTablet or Vista...

Страница 2: ...and enter the PIN provided by your care team default PIN is 1234 Section 1 Setting Up VistaTablet or VistaPhone Status Bar Status Bar Home Key Menu Menu Device Name Device Name VistaTablet Charging P...

Страница 3: ...the device will sound a warning tone and then turn off If this occurs charge the device before turning it on again Checking VitalPatch Hours Each VitalPatch will last up to 5 days 120 hours or 7 days...

Страница 4: ...oval pad underneath c Dispose the VitalPatch according to local laws for routine electronic waste d Clean skin with an alcohol pad and allow to dry Section 4 Removing Previous VitalPatch Sweep Adhesiv...

Страница 5: ...ace two fingers below the jugular notch dip at the base of the neck b Position diagonally with the three lines angled downward away from the midline VitalConnect Midline LEFT Side Left Hand Section 6...

Страница 6: ...back This creates a tab that will help you remove the top liner later See figure to the right 3 Place the Adhesive Overlay Over VitalPatch a Position the adhesive overlay over the VitalPatch with the...

Страница 7: ...Patch ID and then enter the last 6 characters of the Bluetooth ID 3 Choose Pacemaker Status After connecting the pacemaker status screen may appear The screen will appear only once per patient will no...

Страница 8: ...model If the Elevated option is not available ensure the patient is in the Standing Sitting or Supine position when calibrating Instruct patient to remain still in their current body position and tap...

Страница 9: ...lPatch Report Symptom Device Name Organization VitalPatch ID 0087ea has 94 hours remaining 92 5 45 PM Take SPO2 Take BP Take Weight Streaming Mode 98 Berry BM1000 00 01 0C 82 BF 6E Oxygen Saturation H...

Страница 10: ...contents will vary based on device configuration See figures to the right Disconnect Patch Use when monitoring is complete or to pair a new VitalPatch This option maintains the pairing with third part...

Страница 11: ...nt Discharge patient from the monitoring system Cancel Section 10 When Monitoring is Complete Tap Disconnect Patch Choose a Disconnect Option Section 11 Reporting a Symptom Section 12 Making a Video C...

Страница 12: ...e figure below Tap Calibrate to begin the posture recalibration process See Step 5 and 6 beginning on Page 7 If you do not recalibrate VistaPoint will not provide you with respiration rate or posture...

Страница 13: ...itionally if the displayed posture does not match the actual posture you must recalibrate the posture To do this select Posture Settings from the VitalPatch menu and then tap Calibrate Message Fall De...

Страница 14: ...ion Note If you have a pacemaker VistaPoint will not display ECG heart rate or respiration rate Menu Calibrate Posture Calibrate Temperature Flash Patch LED Disconnect Patch Menu Calibrate Posture Cal...

Страница 15: ...tton on VistaPhone and then Take BP 3 Tap Manual Entry if not already selected 4 Tap the first entry field Use the on screen keyboard to enter the two measured values and then tap Enter If you make a...

Страница 16: ...until the display turns off d After confirming operation proceed to Step 2 on page 17 for pairing Setting Up the BM2000A Wrist Refer to the BM2000A Operator s Manual for complete instructions See fig...

Страница 17: ...Saturation tile or Take a Measurement on VistaPhone and then Take SpO2 c Tap Use Bluetooth and then tap Pair New d Tap Berry BM2000A BM1000 as the model e Scanning for Devices will display Tap the Bl...

Страница 18: ...displayed in the upper right of the display no additional action is necessary go to Step 2 below c If Bluetooth is not ON press and hold the Touchpad until the menu screen is displayed Press and hold...

Страница 19: ...n VistaPhone and then Take SpO2 c Tap Use Bluetooth and tap Pair New d Tap Masimo MightySat as the model e Scanning for Devices will display Tap the Bluetooth ID of the Masimo MightySat when it appear...

Страница 20: ...our care team 1 Find the Bluetooth ID Optional Note This step is optional and is only necessary when using the Model 3150 in an environment where multiple similar devices are being used such as a hosp...

Страница 21: ...ings will continue to be transmitted to the dashboard as long as the Nonin 3150 is operational and the sensor is applied to the finger If the dashboard displays Finger Removed check finger is inserted...

Страница 22: ...will be measured and you will be returned to the dashboard If you see a message asking you to confirm the device pairing tap Confirm d Pairing is now complete Remove Model 3230 from finger 3 Take a Re...

Страница 23: ...us step b Tap Take BP in the Blood Pressure tile or Take a Measurement on VistaPhone and then Take BP c Tap Use Bluetooth d Tap Pair New e Tap Model UA 651 as the model f Scanning for Devices will dis...

Страница 24: ...ted c Tap Take Reading or Take a Measurement on VistaPhone and then Take Weight d Step on the scale Step off the scale when it beeps and displays Step Off VistaPoint displays Please wait connecting wh...

Страница 25: ...on 1 Tap the Menu button and then tap Applications menu item Select the Wi Fi Center application this application will not be displayed if your tablet uses a cellular connection 2 If Wi Fi is OFF tap...

Страница 26: ...ap the Volume Down button on the side of the VistaPhone The device will restart normally Bluetooth Connection Issues If you have problems connecting to or disconnecting from the VitalPatch the followi...

Отзывы: