Instruction manual
VitalFlex Elite LTC Bed
Last update:03.16.2021
Page 11 of 34
4.4
Cleaning and disinfection
Before cleaning and disinfection, the mains plug must be disconnected from the mains and securely
hung up. The plug for the handset and the motors, which are plugged into the control at the lying
surface drive, must be plugged in. This is necessary so that no water can penetrate the control
unit.
Do not immerse the electrical components in water, but only wipe them off with a damp cloth.
The electrical components must not be sprayed with a high-pressure cleaner or water jet. Only
wipe disinfection is permitted.
To avoid skin irritation, always wear liquid-impermeable gloves during cleaning and disinfection
work.
Attention: When spray disinfecting with alcohol-containing agents, there is a risk of
explosion and fire when used over large areas.
4.5
Maintenance and repair
Maintenance measures (inspection and maintenance) and maintenance (repair) may only be carried
out by persons who have at least read the safety regulations and followed these operating
instructions.
To detect possible defects in time and to ensure safe use, a technical check (visual and functional
check) must be carried out by qualified personnel at least once a year according to the
maintenance schedule (see chapter 10.2) after a longer period of inactivity and before each reuse.
If the tests reveal errors, damage or defects, the VitalFlex Elite may no longer be operated.
Maintenance of the VitalFlex Elite must be carried out by qualified personnel.
Only original spare parts and accessories of the manufacturer may be used, otherwise all warranty
and product responsibility are excluded.
The 9V block battery is the energy storage device for electrical emergency lowering in the event of
a power failure. The energy storage is sufficient for max. one emergency lowering and must then
be replaced. If the expiry date of the batteries has exceeded, they must also be replaced
immediately. As batteries are self-discharging, it is recommended to replace them every two years
if they are not used. Make sure that this is an alkaline manganese battery of type 6LR61 and that
only this type may be used. Empty batteries must be disposed of in an environmentally friendly
manner.