Instruction manual
VitalFlex Elite LTC Bed
Last update:03.16.2021
Page 27 of 34
10.3
Check of first-error safety by means of integrated locking function in the
handset
Proceed as follows to check the safety device:
The 4 switching positions are test settings which are only used for safety checks as part of the
annual inspection or after repair or before each re-use of the bed.
Move the lock to switching position
I
or
II
: First move all bed adjustments to a slightly raised
position.
Blocking function 1:
Functional test 1
Blocking function 1:
Functional test 2
Blocking function 2:
Functional test 1
Blocking function 2:
Functional test 2
Set the switch positions to test position Locking function 1 or 2 (|): No electrical adjustments must
be possible when the adjustment keys on the handset are pressed.