Appendix
VITA Easyshade Advance 4.0 cannot
be turned on.
VITA Easyshade Advance 4.0 cannot
be calibrated or repeatedly
prompts to calibrate again.
Measurements cannot be made after
switching on the VITA Easyshade
Advance 4.0.
A measurement is made and
VITA Easyshade Advance 4.0
displays "---" in place of a
VITA classical A1–D4 shade or
VITA SYSTEM 3D-MASTER shade.
Inaccurate shades are always
displayed when testing on shade tabs.
A letter (S or W) is displayed
to the right of the Toolbar icon.
Charge the battery as described in section 1.3.
Make sure that the probe tip is correctly placed into the calibration block
holder and held at an angle of 90° and is not lifted before hearing the "beep",
which indicates completion of calibration. Reposition the probe tip as needed.
Make sure that the IR ports of the base unit and the hand piece are not blocked
(see section 1.6.1).
The VITA Easyshade Advance 4.0 first has to be calibrated after switching on
(see section 1.6.). Only then is a measurement possible in the desired measure-
ment mode (see section 4).
Either the color is outside the known range of dental shades,
the probe tip was not correctly positioned or was lifted or moved
during the measurement. Measure again; be careful to use proper
placement technique.
Make sure that Easyshade is in Shade Tab mode. Make sure that the
calibration block and the probe tip are clean; then re-calibrate the unit.
Use only approved VITA Easyshade Advance 4.0 anti-infection covers.
Repeat the measurement, assuring that the probe tip is touching
the surface at 90° in the middle of the shade tab and that the probe tip
is not moving during the measurement. Compare the serial number
to make sure that the correct calibration block for VITA Easyshade Advance 4.0
is installed. Make sure that only VITA Linearguide or Toothguide
3D-MASTER and VITA classical A1–D4 shade tabs are used.
This indicates a temporary change to the instrument, due to environmental
factors, such as static discharge or disconnecting the battery.
Turn the unit off and on again; the letter will not be displayed.
Problem
Then ...
12.3 Troubleshooting