19
28
29
30
VITA Easyshade
®
Advance 4.0
26
27
Modes of operation
4.4 Tooth area measurement on the natural tooth
To measure the cervical, middle and incisal shades of a tooth, highlight
and select the Tooth area measurement icon on the Measurement menu.
Measurement sequence: cervical, middle and incisal.
To begin (empty circle), place the probe on the cervical area of the tooth and
press the measurement switch. Two rapid "beeps" will indicate a successful
measurement and "OK" will be displayed.
Proceed with the measurements of the middle and incisal areas.
After successful measurement of all areas, the VITA classical A1–D4
and VITA SYSTEM 3D-MASTER shade results will be displayed.
If an error occurs during measurement, a long "beep" will sound
and "X" will be displayed in the measured tooth area (not shown).
Repeat the measurement of the area until a successful measurement
result is obtained.
4.3 Averaged measurement on the natural tooth
With the averaged measurement up to 30 measurements can be made on a
natural tooth. The average of the measurement results is displayed as a VITA
classical A1–D4 or VITA SYSTEM 3D-MASTER shade.
The number of averaged measurements is displayed.