The system will now measure the complete length of the rail and it will hit the laser
target plate at different positions.
If you get errors saying the laser target plate was not hit, please check you rail is
straight and that there is no direct strong light hitting the target plate. A sensor is
located inside the target plate ensuring the laser always hits the plate, this too
ensure that correct distance is always measured from the end of the rail.
Strong light on the laser target plate (e.g., from general lighting), must be blocked
with a plate or a dark curtain.
When the rail distance setup is completed, the system will get ready to measure the
sensor laser distance in combination with the rail laser distance to be stored in the
finished calibration. The window below will ask you if you would like to make a
manual pre-check that sensor laser hits
the “Setup laser target plate” as shown
below.
If you click the large button on top of the rail image will you be able to move the
sensor manually using the window shown below.