Please ensure that the laser hits the center of the target plate on the injector by
adjusting the screws next to laser measurement device as shown below.
Also make sure that the “Setup laser target plate” is mounted on the LabSpion with
the front being at the center of rotation of the LabSpion.
The
reason for the “Setup laser target plate” being extended outwards by 10
cm is so
the software can easily ensure that the laser always hits the plate during the setup
process. Otherwise, a significant error of at least 10 cm would be detected by the
software. The “Setup laser target plate” is only used during the setup and is not
needed again after the setup procedure is completed, unless the system is moved.
The sensor must be adjusted in height so that the laser beam is completely
horizontal and points to the center of
the “Setup laser target plate”
as shown below.