Dosing system eco-CONTROL EC200 DUO
Operating and maintenance instructions
Version 1.4
Copyright ©
• Press the
START
key, the calibration starts, the 2K-dispenser is switched on and
conveys the quantity specified by the system. This process can be repeated as often
as desired. e.g., for comparative measurements.
• The ACTUAL quantity that goes into the calibration vessel is determined in the
previously set measuring unit. See also 5.10.4, Dosing quantity.
• Input and confirm the ACTUAL quantity that has been determined.
• Select the OK button and confirm, the next dialog to save the calibration value
appears (display 17).
• Accept or discard the suggested program slot, confirm this,
and the system has been calibrated.
Displ
ay 17
5.9.2
Calibration manual
The
Calibration manual
function provides a simplified option to recalibrate the system
without having to run through the entire process in the
Calibration auto
menu.
This can be helpful, for example, after changing the stator, in the event of batch-
related product variations, etc., whereby the set dosing value is not to be changed
(QA documents, etc.).
Caution:
Setting the flow rate in the
Calibration manual
menu also changes the
dosing results of all the dosing programs by a linear factor.
• Press the
PRG
key,
system selection menu
, display 2
appears.
• Select
Calibration manual
.
• Select motor
A
or
B
, menu
Calibration manual
, display 40
appears.
Displa
y 2
• Confirm the flow rate and set the desired value with the
NAVI wheel
• Press the
NAVI-wheel
key,
system selection menu
, display
2 appears.
Displ
ay 40
Save with OK
Esc
Save calibration value
at prog. no.:
07
OK
Calibration auto
Calibration auto
Calibration manuell
Service
Flush manually A
Flush manually B
Administration
Programs
10 47h
Fr 18.03.11
Accept with OK
Esc
Calibration manual
OK
Degree of dosing qty
Adjustment range 90 ..110%
100.0 %
Содержание PreeFlow eco-CONTROL EC200 DUO
Страница 2: ...Item number 20202 print 21 2 12 ...
Страница 39: ...Dosing system eco CONTROL EC200 DUO Operating and maintenance instructions Copyright Version 1 4 39 43 ...
Страница 40: ...Dosing system eco CONTROL EC200 DUO Operating and maintenance instructions 40 43 Version 1 4 Copyright ...
Страница 41: ...Dosing system eco CONTROL EC200 DUO Operating and maintenance instructions Copyright Version 1 4 41 43 ...
Страница 44: ......