background image

Installation

8

4.

Make sure you allow for a “drip-loop” on the power cord to prevent any water from potentially entering the controller. 
Refer to 

Figure 3

.

5.

Install the UV lamp. Refer to 

Section 4.1

.

6.

Install the cartridges. Refer to

 

Section 4.3

.

7.

When all plumbing connections are made, slowly turn on the water supply and check for leaks. The most likely cause 
for leaks is from o-ring seals. In case of a leak at the chamber, shut water off, remove the retaining nut, wipe the o-ring 
and threads clean and re-install. In case of a leak at the filters, remove the sump, wipe the o-ring and threads clean, 
ensure the o-ring is properly seated, then reinstall.

8.

Once it is determined that there are no leaks, plug the system into the ground fault interrupter, and check controller to 
ensure the system is operating properly.The controller is designed to detect both power to the system and UV lamp 
illumination. 

Note: DO NOT look directly at the glowing UV lamp. 

9.

Allow the water to run for a few minutes to clear any air or dust that may be in the chamber.

Note: When there is no flow, the water in the cell will become warm, as the UV lamp is always on. To remedy this, run a 

cold water tap anywhere in the house for a minute to flush out the warm water. 

3.2 Disinfection Procedure

UV disinfection is a physical process and does not add any potentially harmful chemicals to the water. As UV does not 
provide a disinfection residual, it is imperative that the entire distribution system located after the UV be chemically 
disinfected to ensure that the water is free from any bacteriological contaminants. The disinfection process must be 
performed immediately after the UV unit is installed and repeated thereafter whenever the UV is shut down for service, 
without power, or inoperative for any reason. The procedure for sanitizing the plumbing system is readily accomplished as 
follows:

1

•   

Ensure the controller is 
plugged in for entire 
disinfection process.

2

•   

Shut off the water supply.

•   

Close each faucet.

3

•   

Press the pressure button to 

release the pressure from the 
cartridges.

4

1

2

•   

Remove sump housing using 

sump wrench.

5

      Household 5.25%
       Bleach Solution

BLEACH

x2

2

1

•   

Remove cartridge and pour 2 

cups of household bleach 
solution into the sump 
housing.

Note: DO NOT use Hydrogen 
Peroxide.

6

1

2

•   

Connect only the sump 
housing to the unit.

7

•   

Turn on water supply.

•   

Allow water to fill the 
chamber.

8

0

30

15

45

30 mins

•   

Turn on the cold water supply 

followed by hot water (if 
available) until you smell the 
bleach.

•   

Close all faucets and allow 
bleach to settle in the water 
lines for 30 minutes.

Содержание VH200-F10

Страница 1: ...tario Canada N1L 1R1 t 1 519 763 1032 tf 1 800 265 7246 US and Canada only t 31 73 747 0144 Europe only f 1 519 763 5069 e mail info viqua com www viqua com Models VH200 F10 VH410 F20 Owner s Manual P...

Страница 2: ...y should you insist on genuine factory supplied VIQUA replacement lamps Use of widely available non genuine replacement lamps has been shown to damage the control module of VIQUA UV disinfection equip...

Страница 3: ...started Warning This product may contain chemicals known to the State of California to cause cancer and birth defects or other reproductive harm DANGER Failure to follow these instructions will resul...

Страница 4: ...o clean up a broken lamp as this may scatter the spilled mercury Obey local regulations and guidelines for the removal and disposal of mercury waste NOTICE The UV lamp inside the disinfection system i...

Страница 5: ...tems 5 IEC Replacement Power Cords For VIQUA ICE Controller Sold Separately 260010 NORTH AMERICAN NEMA 5 15P 3 PRONG GROUNDED 602637 CONTINENTAL EUROPEAN CEE 7 7 2 PIN WITH GROUND SCHUKO 260012 UK VER...

Страница 6: ...shows the installation of a typical drinking water system and the related components that may be used for the installation The use of a bypass assembly is recommended in case the system requires off...

Страница 7: ...e system to the wall with appropriate lag bolts not supplied through the two mounting holes located on the metal bracket The use of a flow restrictor device is recommended when installing your system...

Страница 8: ...e in the house for a minute to flush out the warm water 3 2 Disinfection Procedure UV disinfection is a physical process and does not add any potentially harmful chemicals to the water As UV does not...

Страница 9: ...e above disinfection procedure will result in a massive chlorine residual far in excess of the 0 5 to 1 0 mg L typically present in municipally chlorinated water and of a magnitude consistent with the...

Страница 10: ...or UV lamp disposal Do not use water during replacement of UV lamp 1 Shut off the water line to chamber and release system pressure before servicing 2 0 30 15 45 10 mins Disconnect main power source a...

Страница 11: ...quartz sleeve Remove the quartz sleeve 4 Clean the quartz sleeve with a cloth soaked in CLR vinegar or some other mild acid and then rinse with water Note If sleeve cannot be cleaned completely or it...

Страница 12: ...a small bucket under the system to catch any water Procedure 1 Shut off the water flow to the unit depress pressure relief button on top of the filter head to relieve pressure in the filter Refer to...

Страница 13: ...hem P N OR40 50 5 Ensure o rings are liberally coated with a silicone based lubricant 6 Install the new cartridge in the reverse procedure as stated above turning the collar clockwise until the sump i...

Страница 14: ...tting Lamp Life Refer to Section 4 1 Note Even though the alarm on the system can be deferred for a period of time it is important to address each and every alarm condition as they are indicating that...

Страница 15: ...ng seal on retaining nut Ensure O ring is in place check for cuts or abrasions clean O ring moisten with water lubricant and re install replace if necessary Refer to Section 2 for part number Condensa...

Страница 16: ...Manufacturer s Dose Flow Chart 16 Section 7 Manufacturer s Dose Flow Chart Not performance tested or certified by NSF Figure 9 Dose Flow Chart VH200 F10 VH410 F20 VIQUA Standard 32...

Страница 17: ...lbs 13 2 kg 29 lbs Electrical Voltage 100 240 V 50 60 Hz 100 240 V 50 60 Hz Max Current 1 5 Amp 1 5 Amp Power Consumption 35 W 60W Lamp Watts 27 W 45W Maximum Operating Pressure 100 psi 689 kPa 100 p...

Страница 18: ...Hardware Components VIQUA warrants the electrical controller and hardware components to be free from defects in material and workmanship for a period of three 3 years from the date of purchase During...

Страница 19: ...425 Clair Rd W Guelph Ontario Canada N1L 1R1 t 1 519 763 1032 tf 1 800 265 7246 US and Canada only t 31 73 747 0144 Europe only f 1 519 763 5069 e mail info viqua com www viqua com...

Отзывы: