17
Rev. 1 20.12.2019
EN
Temperature measurement ATC – MTC
ATC
: The direct measurement of the sample temperature for all parameters is carried out through the
NTC 30KΩ probe, which can be either integrated into the sensor (electrode and / or cell) or external.
MTC
: If no temperature probe is connected, the value must be changed manually:
keep pressed
or
until the value starts to flash; then adjust it by continuing to use the
directional keys; then press
to confirm.
pH Parameter
pH 7 Vio; PC 7 Vio
On this series of devices, it is possible to use pH sensors with integrated temperature probe or to connect
two different sensors. Connect the pH electrode to the BNC type connector marked in green. Instead,
connect the temperature probe to the RCA / CINCH Temp connector always marked with a green
background.
The instrument is also able to recognize the DHS sensor, an innovative electrode capable of storing
calibration data and then being able to be used immediately on any enabled instrument.
pH parameter Setup
In measurement mode press
to access the SETUP menu.
By pressing the button
access the pH SETTINGS P1.0 menu
Move with the keys
and
to select the sub-menu you want to access.
The chart below shows the setup menu structure for the pH parameter, for each program the options that
the user can choose, and the default value are shown:
Composition of the setup Menu for pH parameter
Program
Description
Options
Factory Default Settings
P1.1
CAL BUFFER SELECT
USA – NIST – Custom
USA
P1.2
SELECT RESOLUTION
0.1 – 0.01
0.01
P1.3
STABILITY CRITERIA
LOW – MEDIUM - HIGH
MED (nor)
P1.6
CALIBRATION DATA
-
-
P1.8
RESET SETTINGS
YES – NO
NO
P1.9
TEMPERATURE CAL
YES – NO
-
P1.1 pH buffer selection
Access this setup to select the buffer family with which to perform the electrode calibration.
Calibration lines from 1 to 3 points can be made.