VMM-T1 User Manual
39
* THIS IS A DISCONTINUED PRODUCT*
Name
Description
Position
Function
Slide Switch
Lower
Upper
TRIGGER INPUT
A
active-low
active-high
PULSE INPUT
B
active-low
active-high
RESET INPUT
C
active-low
active-high
GATE INPUT
D
active-low
active-high
PULSE ENERGY
E
25mm & smaller 35mm & larger
Power Requirements
115 VAC ±5%, 60 Hz or 230 VAC ±5%, 50
Hz., 60 watts
Selected manually by rotating the fuse holder
of the AC power entry module. See "
Voltage
Change
" instructions.
Fuse Requirements
Two 0.5 Amp, “T” slow blow, 3AG for 115 VAC
line
Two 0.25 Amp, “T” slow blow, 3AG for 230
VAC line
One 0.6 Amp, “T”, slow blow, 5 x 20 mm for
Shutter actuator coil
See "LINE FUSE REPLACEMENT"
instructions for AC power fuse change
Operating Temperature
+5° C to +40° C (+41° F to +104° F)
Storage Temperature
-20º C to +55ºC (-4° F to +131° F)
Size (HWD)
2.638" x 5.354" x 8.186" (67.0 x 136.0 x 207.9
mm)
Weight
3.50 lbs (1.59 kg)
Relative Humidity
80% at +31º C (+88 º F) ambient decreasing
linearly to 50% at +40 º C (104 ºF)
Pollution
Degree 2
Overvoltage
Category II