7
MISTRAL EVOLUTION HYBRID
USER MANUAL
www.vicidbs.com
Version 1.03.0000
5.3 Connections
Power Supply: C-20
Outlet N2 with Valve: G1/4” female
Outlet Water Purge: Ø10mm tube
Relay Alarm: 2 pos universal Mate-N-Lock
connector female
5.4 Power Supply Settings
The current power supply setting of the device is indicated on the ID Sticker on the back of the device. The
Mistral Evolution standard setting is 210-230V/50-60Hz.
If other voltages are requested, this should be done when ordering. In this case a transformer will be
installed to transform internally the requested voltage to 110-115 V/50-60Hz. The list of voltage and
frequencies settings possible are bellow. In any case the voltage of the power supplied must stay at
+/- 10% of the nominal value.
VOLTAGE SETTING
FREQUENCY
FUSE F3 (5X20 OR 6.3 x 32 mm)
COUNTRY
110-115 VAC
AC50-60 Hz
20 A
Japan/U.S.A.
220-240 VAC
AC50-60 Hz
10 A
Korea/Europe
Please see service manual for the connection of the transformer. General earth connection is done via power
supply main cable.
5.5 Transport and Storage Conditions
Temperature
Minimum -10ºC (14ºF), Maximum 45ºC (113ºF)
Humidity
Inferior to 90% Rh, non-condensing
Shocks and vibrations during transport
Shocks <1g; Vibrations <0.2g/5Hz
Instruments position
Horizontal