10
MISTRAL EVOLUTION HYBRID
USER MANUAL
www.vicidbs.com
Version 1.03.0000
• Read the Operations Manual carefully.
• Do not place the Mistral Evolution on a sloping area, use the wheel breaks.
• Make sure the voltage setting is correct
• The power supply main cable connector is the disconnect device in case of emer-
gency, it must be easily accessible at anytime.
7. STARTING THE GENERATOR
Connect the Mistral Evolution to the power line.
1.
Close the On/Off (O/I) valve placed at the back side of the device
2.
Connect the outlet to your application
3.
If you use an intermediate buffer tank, please look at the specific chapter. Connect the outlet water
condensate of the Mistral Evolution to an external water tank
4.
Turn the generator on (switch in back).
5.
Press the MENU button and scroll until you get the PARAMETER menu. Press the SELECT button.
6.
Go into each sub menu and make your selections. The different possible setting are explained in the
specific chapter
7.
Once done, press EXIT two times to get back at the screen displayed in point 5.
8.
Press START button
9.
The compressors will start and the pressure will increase to 8 barg (this pressure is set at 8 barg in the
factory). If you need a different pressure setting, open the service box at the back side of the device and
make your adjustment. Turn right to increase the pressure or turn left to decrease it.
10.
Once the pressure set, open the ON/OFF valve at the back side of the device.
11.
Warm-up the generator for several hours to reach the correct Nitrogen purity.