3. SAFETY
3.1 Personal Safety Instructions
When using your Mistral Evolution Hybrid, basic safety precautions should always be followed, including the
following:
• The quantity of dust in the ambient Air can be very different from one place to another.
We recommend to replace the dust filters every 4000H. In some dusty areas, this
maintenance need to be performed more frequently.
•
Clogged dust filter may damage the device or give a bad Nitrogen quality.
• The power supply main cable connector is the disconnect device in case of
emergency, it must be easily accessible at any time.
•
If the Mistral Evolution is used in a way not specified by VICI DBS, the protection
provided by the equipment could be compromised.
3.2 Instrument Safety Instructions
• The Mistral Evolution Hybrid is a Nitrogen Generator. Do not let it run with the outlet
open to the air in a low volume room.
• The increase of the Nitrogen (and decrease of oxygen) may cause human disease.
It is the responsibility of the user to check if the room is appropriate or if an O2
measurement is necessary.
• The Mistral Evolution Hybrid rejects Oxygen as waste.
• Oxygen ratio in the air could change around the device in a low volume or not
sufficient ventilated room.
•
If fuel gases are present around, their explosive or flammable limits could be
affected. It is the responsibility of the user to check if the room is appropriate or
if an O2 measurement is necessary.