background image

Copyright © VibroSystM Inc., 2010

31

LIN-300 Series Modules_Installation - Rev.4

O

4. INSTALLATION OF LIN-300 SERIES MODULES

4.1 Preliminary considerations

LIN-300 Series modules are signal condi-
tioning units which, when supplied with 24VDC
and paired with matching sensors, deliver a 4
to 20 mA linearized raw output signal repre-
senting a distance which can be used by
AGMS and ZOOM systems, ZPU-5000, PCU-
5000, PCU-100 Programmable Monitor, or
third-party instrumentation.

The sensor is connected to the LIN-300
module through a triaxial cable with an SMA
connector, while power input and signal output
share an extension cable connected to the
LIN-300 M12 socket.

Figure 23 : LIN-300 Series module

LIN-300 modules are usually installed in pairs on a mounting plate enclosed in a wall-mounted protection
box. 

+24 V

DC

 power input is required to power up the LIN-300 modules. Power can be provided from an

external source, such as from an XPSP-224P External Power Supply panel.

The extension cable can be assembled on site, using 4-conductor shielded cable and an M12 connector,
or a standard molded M12 cable assembled can be used. The standard cable offered by VibroSystM is a
molded M12 cable, 30m (100 ft.) long, and is rated for a maximum operating temperature of 80

°

C (176

°

F).

Since this cable is not calibrated, it can be cut. 

Installation of the LIN protection box, protective conduits, and triaxial cables should be completed before
proceeding with the installation of the M12 extension cable, and connection of the LIN-300 modules.

A flexible conduit must be installed between the LIN-300 protection box and the location of the power
source and data-receiving instrumentation. This conduit, attached with liquidtight connectors and
supported by conduit clamps, will protect the M12 extension cable.

Cable length from LIN-300 signal conditioner to instrument should not exceed 300m (1000 ft.) for the 4-
20 mA output signal. 

A connection that has not been correctly tightened or has become loose can cause erratic
readings. To ensure correct system operation, all connections must be correctly
tightened.

Содержание LIN-300 Series

Страница 1: ...al P N 9428 25I2A 311 Applies to measuring chain model LIN 304S 6I 0 3 4 3 LIN 306S 6I 0 3 6 3 This manual divides into the following sections Section 1 Overview of the PCS 200ESB Measuring Chain Sect...

Страница 2: ...Copyright VibroSystM Inc 2010 2 PCS 200ESB with LIN 304S LIN 306S_Intsallation_Contents_Rev 1 O...

Страница 3: ...clamping the integral cable 15 2 3 General specifications of the PCS 200ESB sensor 16 3 INSTALLATION OF THE LIN 300 PROTECTION BOXES 3 1 Installation of the 10X6X3 ABS Protection Box 17 3 1 1 Prelimi...

Страница 4: ...nal power source 33 4 2 2 1 Field assembly of an M12 connector onto extension cable 33 4 2 2 2 Connection of the M12 extension cable Power Input and Signal Output 34 4 2 3 Connection of modules in a p...

Страница 5: ...O 1 OVERVIEW OF THE PCS 200ESB MEASURING CHAINS WITH LIN 300 SERIES CONDITIONERS Figure 1 Typical PCS 200ESB LIN 300 Series measuring chains installation LIN 304S OR LIN 306S LINEARIZATION MODULES IN...

Страница 6: ...ies Signal Linearization Module either a LIN 304S 6I 0 3 4 3 Linearization Module for a 4mm 157 mils detection range b LIN 306S 6I 0 3 6 3 Linearization Module for a 6mm 236mils detection range Option...

Страница 7: ...vice personnel Safety and electrical symbols that appear in this manual and on the material 1 2 Planning and installation requirements Respect the following guidelines when planning your equipment con...

Страница 8: ..._Overview_Rev 2 O 1 3 General specifications of measuring chains Operation Sensor PCS 200ESB Matching linearization modules LIN 304S 6I 0 3 4 3 LIN 306S 6I 0 3 6 3 Output 4 to 20 mA linearized signal...

Страница 9: ...t of the PCS 200ESB sensor must be of conductive semi conductive material The integral cable on the standard PCS 200ESB is protected on its full length by flexible stainless steel cable armor Figure 2...

Страница 10: ...avoid crosstalk special care must be taken while routing the integral cables Two cables should never be inserted in the same conduit Never install integral cables in parallel close to one another Keep...

Страница 11: ...brackets 2 2 Step by step installation of the PCS 200SB sensor The installation of PCS 200ESB normally includes the following steps choosing the sensors location fabrication of mounting brackets inst...

Страница 12: ...brackets Factors to be considered when machining the mounting bracket The mounting bracket must be installed at a perfectly stable location The sensor is 3 4 in diameter with 3 4 16 UNF thread It can...

Страница 13: ...d measure signal output The distance is correct when the conditioner output reads precisely 12 mA Grounding The principle of operation of the PCS 200ESB sensor requires adequate high frequency groundi...

Страница 14: ...th a ohmmeter for an open circuit condition between the mounting bracket and insulated bearing 2 Place a temporary cable between the sensor casing and insulated bearing to create a short circuit path...

Страница 15: ...d cables over the whole distance from PCS 200ESB sensors to LIN 300 Series linearization modules Figure 11 Cable secured with cable clamps Caution To prevent damage to the cable do not force beyond th...

Страница 16: ...range Environmental Temperature range 50 to 200 C 58 to 392 F Humidity Up to 95 non condensing Connection Integral Cable Type Triaxial Length 5 5 m 18 ft minimum Connector Male SMA with ground lug sta...

Страница 17: ...idtight connectors Prepare the openings on the protection box and set the liquidtight connectors in place before fastening the protection box To determine the best location for the protection box surv...

Страница 18: ...anchors 4 bolts M6 1 00 x 16 mm 4 spring lockwashers 1 4 4 flat washers 1 4 4 brackets 4 spring lockwashers 10 32 4 flat washers 10 32 4 screws 10 32 x 1 2 Figure 3 Protection Box Installation kit 3...

Страница 19: ...ace the anti vibration rubber mounts added to the mounting brackets Before deciding on the location for the holes make sure that the liquidtight connectors will not be too close to the LIN conditioner...

Страница 20: ...ing a protection box on the stator or machine casing 1 Drill refer to Figure 6 Mounting holes location and thread into the structure refer to Figure 7 Fastening the ABS protection box on steel 2 Faste...

Страница 21: ...Anchor setting punch for concrete anchors 2 Fasten the ABS protection box as shown refer to Figure 8 Fastening the ABS protection box on concrete Figure 8 Fastening the ABS protection box on concrete...

Страница 22: ...tached with liquidtight connectors Prepare the openings on the protection box and set the liquidtight connectors in place before fastening the protection box Figure 10 Front view of the 14x12x8 protec...

Страница 23: ...4 5 bolts 1 4 20 x 5 8 1 bolt 1 4 20 x 1 1 4 4 concrete anchors 2 copper lugs 1 ground copper wire 5m 2 liquidtight strain relief connector 2 nylon locknuts 3 4 2 rubber adapters for liquidtight strai...

Страница 24: ...embly 2 Fastening the protection box 3 Grounding the protection box 3 2 5 Preparing the holes for the liquidtight connectors and grounding assembly 1 As a precaution before drilling the protection box...

Страница 25: ...er grounding wire To determine the location for the holes keep in mind the components on the mounting panel Make sure the compo nents will not get in the way of the connectors once the mounting panel...

Страница 26: ...strain relief connectors Refer to Figure 16 Mounting the liquidtight strain relief connectors 6 Install the connectors 7 Reinsert the mounting panel Figure 16 Mounting the liquidtight strain relief c...

Страница 27: ...n box on steel 2 Fasten the protection box as shown Figure 18 Fastening the protection box on steel Mounting a protection box on concrete 1 Drill into the concrete wall refer to Figure 17 Mounting hol...

Страница 28: ...operation integrity For best grounding provide the shortest path possible between the protection box and the grounded structure 1 Drill and tap a hole in a grounded structure Make it as close as possi...

Страница 29: ...tection Boxes_Installation Rev 0 O 3 Cut the copper wire and fasten it to the outside of the protection box as shown in Figure 22 Fastening the grounding wires to the protection box Figure 22 Fastenin...

Страница 30: ...Copyright VibroSystM Inc 2010 30 LIN 300 Protection Boxes_Installation Rev 0 O...

Страница 31: ...ernal Power Supply panel The extension cable can be assembled on site using 4 conductor shielded cable and an M12 connector or a standard molded M12 cable assembled can be used The standard cable offe...

Страница 32: ...ls needed fish tape to pull M12 extension cable inside the protective conduit wire stripper assorted miniature flat screwdriver 4 2 Step by step installation 4 2 1 Connection of the SMA triaxial cable...

Страница 33: ...ding braid back 4 Remove foil and nylon thread under the shielding braid 5 Strip 5 mm off the insulation of the wires Install the wires to the female connector the pin out table push together and fast...

Страница 34: ...quipped with an internal ratchet element preventing unintentional loosening of the compression nut M12 connectors with this feature produce a clicking sound when turning the compression nut When the r...

Страница 35: ...be connected the power source cable to the Power In AC DC terminals the signal output extension cable to the Out1 and Out2 terminals Nominal input for miniature power supply AC Input voltage 100VAC 2...

Страница 36: ...eral instruments may be connected to a circuit protected by a circuit breaker but each instrument must have its own disconnect device A power source cable must be routed to the LIN 300 protection box...

Страница 37: ...conductor may be fitted with a crimped ferrule if desired 4 Connect each pair of wires to the proper terminals Figure 30 Signal output cable connected to OUT1 and OUT2 terminal blocks Note A pair of t...

Страница 38: ...he status of the LIN 300 Series signal conditioner The sequence at startup is as follows 1 an orange flash followed by 2 a green flash followed by 3 a green or red pulsation A green pulsation one brie...

Страница 39: ...to 6 3 mm 12 to 248 mils Output 4 to 20 mA Sensitivity 2 667 mA mm 67 8 A mil Accuracy 3 of reading at mid range Repeatability 0 3 of reading at mid range Interchangeability 3 of reading at mid range...

Страница 40: ...Copyright VibroSystM Inc 2010 40 LIN 304S and LIN 306S Specifications_Rev 1 O...

Отзывы: