VibrationVIEW Help
247
© 1995-2005 Vibration Research Corporation. All rights reserved.
7.6.31 Maximum G RMS (chN)
Exceeded RMS Acceleration Limit (
Field Data Replicator Stop Code
)
Maximum Control (Ch1) G rms
. This code indicates that the maximum allowable control Grms was
reached while running your test. Perhaps the accelerometer current source is off or your amplifier is
not turned on. You may need more Gs. Modify the appropriate parameter found in the
dialog box.
Maximum Reference (Ch2) G rms
. This code indicates that the maximum allowable control Grms was
reached while running your test. You may have a large spike in your data or the channel two sensitivity
is not set correctly. You may need more Gs. Modify the appropriate parameter found in the
dialog box.
7.6.32 Maximum Run System Gain
Maximum Run System Gain (
Field Data Replicator Stop Code
)
This code indicates that the ratio between the output voltage and the measured acceleration exceeds
the maximum value allowed while running the test. Perhaps the accelerometer is disconnected or your
amplifier is not turned on. You may need to adjust the running Max System Gain parameter, found in
the
dialog box.
7.6.33 Maximum Start System Gain Abort
Maximum Start System Gain Abort (
Field Data Replicator Stop Code
)
This code indicates that the ratio between the output voltage and the measured acceleration exceeds
the maximum value allowed while starting the test. Perhaps the accelerometer is disconnected, or
your amplifier is not turned on. If necessary, you may need to adjust the maximum system gain
parameter, found in the
dialog box.
7.6.34 Max Run Voltage
Max Run Voltage (
Field Data Replicator Stop Code
)
This code indicates that the maximum allowable voltage was reached when running your test. Perhaps
the accelerometer is disconnected or your amplifier is not turned on. You may need more voltage.
Modify the appropriate parameter found in the
dialog box.
7.6.35 Maximum Start System Gain
Maximum Start System Gain (
Field Data Replicator Stop Code
)
This code indicates that the ratio between the output voltage and the measured acceleration exceeds
the maximum value allowed while starting the test. Perhaps the accelerometer is disconnected or your
amplifier is not turned on. You may need to adjust the starting Max System Gain parameter, found in
the
dialog box.
7.6.36 Max Run System Gain Abort
Max Run System Gain Abort (
Field Data Replicator Stop Code
)
This code indicates that the ratio between the output voltage and the measured acceleration exceeds
the maximum value allowed while starting the test. Perhaps the accelerometer is disconnected, or
your amplifier is not turned on. If necessary, you may need to adjust the maximum system gain
parameter, found in the
dialog box.
195
195
195
195
195
195
195
Содержание VibrationVIEW
Страница 15: ...Document Revision Part I ...
Страница 17: ...How To Part II ...
Страница 96: ...Menu commands Part III ...
Страница 142: ...Test Definition Part IV ...
Страница 217: ...Control Centers Part V ...
Страница 226: ...VibrationVIEW Help 213 1995 2005 Vibration Research Corporation All rights reserved See Also RecorderVIEW 309 ...
Страница 227: ...Accelerometer Toolbars Part VI ...
Страница 231: ...Stop Codes Part VII ...
Страница 263: ...Typical Views Part VIII ...
Страница 265: ...Typical Views 252 1995 2005 Vibration Research Corporation All rights reserved 8 4 Random Random CRT View ...
Страница 268: ...VibrationVIEW Help 255 1995 2005 Vibration Research Corporation All rights reserved ...
Страница 269: ...Dialog Boxes Part IX ...
Страница 303: ...Windows Elements Part X ...
Страница 309: ...APPENDIX Part XI ...
Страница 311: ...APPENDIX 298 1995 2005 Vibration Research Corporation All rights reserved ...