Chapter 3
Powermeter and VFL
Characteristics of the options
ONA-800 Getting Started Guide
Page 42
22142812, R011
April 2022
Characteristics of the options
The following sections describe the specifications of the Powermeter and VFL options.
Powermeter option
Specifications given for 25
o
C, after 20 minutes stabilization time and after zero setting.
•
Wavelength range: 800 to 1650 in steps of 1 nm
•
Calibrated wavelengths: 850/1310/1490/1550/1625/1650 nm
1
•
Accuracy at calibrated wavelengths: ± 0.2 dB (at -30 dBm)
•
Input power range: -60 dBm to +10 dBm
•
Maximum resolution: 0.01 dB/0.01 nW
•
Measurement range: +5 to -50 dBm (-45 dBm from 800 to 1250 nm)
•
Linearity within the measurement range: ± 0.2 dB
VFL option
•
Wavelength: 650 nm
•
Maximum power: 1.2 mW
•
Length of fiber: up to 5 km
•
Class 2 laser standards: IEC 60825-1 and FDA 21 CFR part 1040.10
1.Specifications guaranteed to the calibrated wavelengths, except for 1650 nm