Viatran DL Series Analog Input Process Meter Instruction Manual
9
NORMAL MODE
REJECTION
Greater than 60 dB at 50/60 Hz
ISOLATION
4 kV input/output-to-power line
500 V input-to-output or output-to-P+ supply
OVERVOLTAGE
CATEGORY
Installation Overvoltage Category II:
Local level with smaller transient overvoltages than
Installation Overvoltage Category III.
ENVIRONMENTAL
Operating temperature range: -40 to 65°C
Storage temperature range: -40 to 85°C
Relative humidity: 0 to 90% non-condensing
CONNECTIONS
Removable screw terminal blocks accept 12 to 22 AWG
wire, RJ45 for external relays, digital I/O, and serial
communication adapters.
ENCLOSURE
1/8 DIN, high impact plastic, UL 94V-0, color: black
MOUNTING
1/8 DIN panel cutout required:
3.622" x 1.772" (92 mm x 45 mm)
Two panel mounting bracket assemblies are provided.
TIGHTENING
TORQUE
Screw terminal connectors: 5 lb-in (0.56 Nm)
OVERALL
DIMENSIONS
4.68" x 2.45" x 5.64" (119 mm x 62 mm x 143 mm)
(W x H x D)
WEIGHT
9.5 oz (269 g)
WARRANTY
3 years parts & labor
Содержание DL-ACB0N
Страница 51: ...Viatran DL Series Analog Input Process Meter Instruction Manual 51 Pump Alternation Control Operation ...
Страница 89: ...Viatran DL Series Analog Input Process Meter Instruction Manual 89 Figure 22 1 8 DIN Panel Cutout Template ...
Страница 90: ...Viatran DL Series Analog Input Process Meter Instruction Manual 90 This Page Intentionally Left Blank ...
Страница 91: ...Viatran DL Series Analog Input Process Meter Instruction Manual 91 NOTES ...