background image

Pub# OM22-iScan EC-OM 

 

6-5 

Procedure #4: Diagnosing and Correcting EC Signal Problems.

 

 
 

 

   

Figure 4-1.

 

 
Electrode Functions-

 

 

Each coulter electrode is part of a pair, and each pair has a distinct function.  
On an iScan: 

Coulters 2 & 3 are the receptors

—they measure voltage drop.  

Coulters 1 & 4 

are the “charged” coulters that inject the current into the soil. If you are getting no (or 

intermittent) readings -- continuity to one of these two coulter-electrodes is likely the cause. 

 
If the continuity ohm test indicates a problem on a channel, you will need to determine where the 
interruption is located.  Listed below are detailed instructions on how to determine exactly where a 
continuity or isolation problem is located: 
 
A. Testing Cable and Wiring continuity: 
 

1.  Once a high resistance reading (poor continuity) on a channel is confirmed, determine whether the 

problem is in the wiring or in the coulter-electrode. To test all cable and wiring, place one ohmmeter 
lead in the Test Box terminal pin for that channel and the other on the corresponding electrode wire 
terminal bolt (Figure 4-2). Repeat process on all electrodes. 

                                   

           

 

Figure 4-2.    
 

2.  If you see <2 ohms on all, the cable is likely ok; test the coulter electrodes as explained in B below. 
 

Electrode #1 
Back side of 
main iScan 
member 

Electrode #2 
Front side of 
main iScan 
member 

Electrode #3 
Top side of 
rear shank 

Electrode #4 
Underneath 
closing 
wheel mount 

Содержание iScan EC-OM

Страница 1: ...em Check Section 5 5 1 Field Operations 5 3 Collecting Quality Data Section 6 Maintenance and Service Procedures 6 1 1 OM signal testing 6 2 2 EC signal testing 6 3 3 Testing electrical continuity 6 5...

Страница 2: ...pon inspection to be defective in materials or workmanship Veris Technologies shall have no obligation under this warranty for the cost of labor down time transportation charges or for the repair or r...

Страница 3: ...and clear when raising or lowering any part of the implement Install all transport locks before transporting or working underneath Detach and store implements in an area where children normally do not...

Страница 4: ...r 9 01 2016 contain an internal Garmin 15x GPS No external GPS is provided on these units however external GPS receivers can be used The controller will automatically give preference to any external G...

Страница 5: ...store it indoors Figure 3 iScan EC OM Figure 4 V Sense Controller Figure 5 V Sense Controller rear Figure 6 V Sense Controller front On Off Turns power to V Sense Controller On Off 3A Fuse OM Comm Se...

Страница 6: ...anel and double click on Region Step 2 Click on Additional Settings the following screen will appear The Decimal symbol needs to be a while the Digit grouping symbol needs to be a The will ensure prop...

Страница 7: ...can EC OM 2 4 Figure 9 Figures 10 Figure 11 Next two license agreements will need to be accepted before continuing The installer will install all necessary components Click Next to continue through in...

Страница 8: ...Pub OM22 iScan EC OM 2 5 Figure 12 Figure 13 Figure 14 Click Next to install the USB drivers Click Next to continue through installation Click Finish to complete the installation of SoilViewer...

Страница 9: ...components of the unit Figure 1 Figure 2 Coulter depth control wheel Turbo coulter EC electrode 1 Air bag Lift point Typical planter mounting location V Sense Controller iScan row unit EC electrode 2...

Страница 10: ...n nuts securely Position bracket centered between any existing implement row units Be careful not to pinch any existing cables or hoses Reroute these as needed Figure 4 Step 3 Lower iScan module onto...

Страница 11: ...Route all cables carefully and secure to prevent damage Figure 8 Step 6 iScan removal Reverse installation process described above After V Sense Controller and four bolts attaching iScan module mounti...

Страница 12: ...ub OM22 iScan EC OM Section 4 SoilViewer and V Sense Controller Set up Select iScan Series Name the data file for sensor data All sensor data is stored in a single file VSENSE RAW Figure 1 Figure 2 4...

Страница 13: ...ignal received The OM EC Comm Light must be green indicating the PC and V Sense controller are communicating properly The EC value must be greater than 1 When all these conditions are met the Saving D...

Страница 14: ...ound but EC readings are positive Check for a moist soil buildup or a pinched cable causing a short between EC electrodes Error EC_NP Row unit is engaged in soil but EC readings are negative Check for...

Страница 15: ...ndow is clean and in good condition see below Place the dark side of the reference block under the window and click continue Then the next message will appear Figure 7 Flip the reference block over to...

Страница 16: ...and the right has about 2500 acres on it Inspect the leading edge shown below as the steel wears the window can chip or crack As this contiues to wear it will eventually need replaced Figure 10 Wearp...

Страница 17: ...roper Operating Depth Begin field operation by lowering implement into soil iScan module must be operated in level position For good electrical conductivity readings all four electrodes must be in dir...

Страница 18: ...hat they are making aggressive soil contact This typically requires a shallow setting If soil slot is not closing properly the rear closing depth wheels are adjustable to increase or decrease the amou...

Страница 19: ...ildup 2 Check angle of implement is system running level 3 Verify optical module is firmly pressing into moist soil at a 1 5 3 3 6 cm depth 4 Check plant residue buildup around iScan runners remove if...

Страница 20: ...t lock iScan into non functioning position Figure 8 Figure 8 Problem iScan module is plugging with soil and or leaving too much of a trench Solutions 1 Reduce down pressure on airbag 2 Lower coulter d...

Страница 21: ...3 that will enable you to quickly check the instrument to ensure that it is functioning properly To perform this test 1 Disconnect the optical power cable from the V Sense Controller 2 Connect the tes...

Страница 22: ...eck the instrument to ensure that it is functioning properly To perform this test 1 Disconnect the signal cable from the signal terminal on the V Sense controller 2 Connect the test load to the signal...

Страница 23: ...tting ohms rather than kilo or mega ohms If unfamiliar with ohming out or resistance testing make a few trial tests before performing the Veris signal test procedure Touching the meter leads together...

Страница 24: ...hromium carbide runner Figure 3 3 Next move to pin number 3 on the test box and touch ohmmeter lead to electrode 3 which is the rear chromium carbide runner with optical module Figure 3 4 Finally move...

Страница 25: ...elow are detailed instructions on how to determine exactly where a continuity or isolation problem is located A Testing Cable and Wiring continuity 1 Once a high resistance reading poor continuity on...

Страница 26: ...th for the EC signal to travel A special hub with a spring plunger presses against the spindle of the coulter serving as a commutator Shown below is a cut away view of the hubcap and plunger assembly...

Страница 27: ...soil bridging across insulators is evident the wires inside the EC signal cable may be shorting Begin by ruling out any problem with the V Sense controller disconnect signal cable from V Sense control...

Страница 28: ...clockwise 4 Depress the spring loaded tip on a hard surface to determine if plunger has adequate tension and can move freely 5 If the plunger will not move freely replace Coat the threads with di elec...

Страница 29: ...grease and install as outlined above 4 Re install hub cap and tighten firmly You may have to re set the plunger to compensate for the reduced length on the newly ground spindle Re adjust as outlined a...

Страница 30: ...bit Verify that your GPS cable is sending GPS data through pin 2 pin 5 is ground and no other pins have signal or power on them The most common issue is hz rate If the GPS has been used for lightbar...

Страница 31: ...e 7 2 Turbo coulter hub Use good quality wheel bearing or lithium grease for lubrication but we suggest that you grease the hub sparingly Over lubricating the hub will result in pre mature seal failur...

Страница 32: ...Procedure 5 7 Reassemble and adjust bearing pre load by fully tightening spindle nut then backing off until you can turn the hub fairly easily with one hand normally this involves backing off 1 2 slot...

Страница 33: ...ates to the runner Figure 9 1 3 To remove wear plate unscrew the hex bolts on the top of the optical sensor as shown in figure 9 2 4 To replace the wear plate ensure the O ring is seated in the O ring...

Страница 34: ...is listed here In this case a USB to serial converter is being used and the port is listed as USB Serial Port COM33 Figure10 1 If the port is not listed here then unplug and replug the USB Serial con...

Страница 35: ...ng at distances of 150 feet and eliminating the need for a physical connection to the PC Follow the directions below to establish this connection Click the Bluetooth icon in the lower right hand corne...

Страница 36: ...Pub OM22 iScan EC OM 6 16 Enter the passcode 1234 then select Next V Sense controller will show Connected and be ready for use Figure 11 4 Figure 11 5...

Отзывы: