60
PROCEdURE 3. CALIBRATE THE PROBE USING THE SCANPOINT SYSTEM
If you do not have ScanPoint
®
with QuickPrint software, you must send your instrument to an authorized
Verathon
®
service center for calibration. Contact Verathon Customer Care for more information.
At minimum, the AMI 9700 must be calibrated every 12 months in order to ensure appropriate results.
Calibrating ensures proper alignment of the instrument’s internal coordinate system. If calibration is not
performed by the prescribed date, the instrument can still be used to take scans but measurements may be
compromised. When calibration is required, a warning appears in the display header.
Figure 23. Calibration Warning
1. Within 10 feet of the Battery Charger/Wireless Hub, place the calibration tank on a flat, nonreflective
surface, and then remove the lid.
2. Pour clean, room‑temperature water into the container, filling to the indicator mark. Ensure that there is a
minimal amount of bubbles in the water.
3. Using the notches to position the spiral‑shaped target correctly, place the target in the container.
4. Replace the lid onto the calibration container.